<BISx Incompatible> is displayed.
Cause : The sensor is not compatible with the monitor configuration.
Solution : Replace the BISx.
<Check BIS Sensor, Perform Sensor Check> is displayed.
Cause : Problem is detected relating to sensor ground element, or sensor is using too much
current.
Solution 1 : Disconnect and examine sensor connection. Clean any contamination.
Solution 2 : Replace the sensor if necessary.
Solution 3 : Replace the patient interface cable.
Solution 4 : Replace the BISx.
<Replace BIS Sensor, Too Many Uses> is displayed.
Cause : Sensor has been connected and disconnected too many times.
Solution : Replace the sensor.
<Replace BIS Sensor, Invalid Sensor> is displayed.
Cause : Sensor is invalid.
Solution : Replace the sensor.
<Sensor Usage > 24hrs.> is displayed.
Cause : Sensor was attached to the system for more than 24 hours.
Solution : Replace the sensor.
<BISx Failure> is displayed.
Cause 1 : Poor connection between the BISx and monitor.
Solution : Follow on-screen instructions. If necessary, turn the power OFF and unplug the power
cable to shut down the monitor completely. Then plug in and re-start the monitor.
Cause 2 : BISx is defective.
Solution : Replace the BISx, then perform "Sensor Check".
Cause 3 : Patient monitor is defective.
Solution : Replace the patient monitor.