-82-
Message
Reason
Remedy
ERROR 11
UNDER RANGE
Detected level of power
meter is too small.
In the case of P2 or P3 mode, the
power of the incident light is under the
detectable level of the power meter.
Check fibers and connection. Set fiber
again if the offset of fiber axis is large.
ERROR 12
GPIB INACTIBE
Power meter is not
active.
In the case of P2 or P3 mode, the
power meter is not active or GP-IB
cord is not connected properly in the
automatic aligning sequence. Check
the power meter is on and GP-IB
connection. Check the address switch
of the power meter is TALK ONLY.
ERROR 13
LEVEL
UNCHANGED
The level of power meter
doesn’t change.
X,Y, PL or PR motor doesn’t move
normally.
Consult the sales agency or factory.
ERROR 14
COVER OPEN
Wind protector is open
during PL or PR motor
moving.
Close the wind protector and press
RESET key.
Содержание FSM-20PMII
Страница 1: ...II ...
Страница 2: ... 1 1 GENERAL 1 GENERAL 1 GENERAL 1 GENERAL ...
Страница 4: ... 3 2 CONSTRUCTION 2 CONSTRUCTION 2 CONSTRUCTION 2 CONSTRUCTION ...
Страница 15: ... 14 4 OPERATION 4 OPERATION 4 OPERATION 4 OPERATION ...
Страница 18: ... 17 Hear is Flowchart page 1 ...
Страница 19: ... 18 Hear is Flowchart page 2 ...
Страница 42: ... 41 5 MENU COMMANDS 5 MENU COMMANDS 5 MENU COMMANDS 5 MENU COMMANDS ...
Страница 52: ... 51 6 DIP SWITCH 6 DIP SWITCH 6 DIP SWITCH 6 DIP SWITCH FUNCTIONS FUNCTIONS FUNCTIONS FUNCTIONS ...
Страница 58: ... 57 7 OHTERS 7 OHTERS 7 OHTERS 7 OHTERS ...
Страница 66: ... 65 8 MAINTENANCE 8 MAINTENANCE 8 MAINTENANCE 8 MAINTENANCE ...
Страница 80: ... 79 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING ...
Страница 87: ... 0 1 0 0 2 3 04 5 6 5 0 7 8 0 2 3 1 0 6 0 0 0 20 9 6 3 0 0 9 3 0 0 20 9 0 0 6 8 0 6 8 0 2 0 20 0 1 6 8 0 1 9 9 ...
Страница 88: ... 5 6 2 0 0 0 2 0 0 0 5 5 3 1 2 1 26 6 20 5 2 0 0 2 0 0 9 A 2 6 3 0 0 0 0 2 3 A 2 6 8 0 0 1 ...