010-021-20 09.2001
Maintenance and Inspection
During maintenance and inspection, strictly observe precautions contained in
"1.2 Operational Precautions"
(page 1-3)
and
"1.3 Safety Precautions" (page 1-7)
in this manual for you to use the FCR XG-1 system
under best conditions.
1 User's Maintenance and Inspection Items
1.1 Daily Inspection
(1)
Inspection Before Use
•
Make sure that the equipment starts up normally.
•
Make sure that the equipment communicates with connected devices normally.
➡
"2.1.1 Starting Up and Shut Down the System" (page 2-2)
(2)
Inspection During Use
•
Make sure that the equipment handles cassettes normally.
➡
"2.2 Routine Operations" (page 2-8)
•
Make sure that the cassette opens / closes and locks normally.
•
Make sure that images are output normally.
If any failures, such as transparent image, etc., result, clean the equipment.
➡
"2.3 Cleaning" (page B-3)
(3)
Inspection After Use
•
Make sure that the power turns OFF normally after the equipment has been shut down.
➡
"2.1.1 Starting Up and Shut Down the System" (page 2-2)
1.2 Periodical Inspection
Using a vacuum cleaner, remove any dirt or dust accumulated in each unit of the equipment once every three
months. Clean then with a slightly moistened soft cloth and wipe off any moisture with a dry cloth.
➡
Figure 1.1 in "1.4 Hardware Configuration" (page 1-8)
No.Unit for Cleaning
No.Unit for Cleaning
No.Unit for Cleaning
1
Main unit cover
3
Cassette inserting unit
5
Periphery of devices
2
Operation panel
4
Air filter
6
CR Console monitor
Z
Be sure to turn the equipment power OFF before cleaning the filter.
2 Maintenance and Inspection Items Assigned to Specified Dealer
For periodical inspection of the equipment and necessary arrangements, consult our official dealer, Fujifilm
Medical Systems representatives of each region.
HHS Label #1
HHS Label #1
EN60825-1:1996
Class
III
b Panel Label
The air filter on the left side of the device can be removed as an assembly, together with the filter cover.
Remove first the cover and then the air filter (plastic) fit in the cover. Clean them using a vacuum cleaner.
For the air filter inside the front panel, first remove the front panel.
The air filter inside the front panel can be removed as an assembly, together with the filter cover.
Remove the cover and then air filter (sponge) fit in the cover. Clean them using a vacuum cleaner.
AVOID DIRECT EXPOSURE TO BEAM
Laser radiation when open
DANGER
HHS Label #1
CAUTION
VORSICHT!
ATTENTION
ATTENZIONE
PRECAUCION
OBSERVERA
LASER RADIATION WHEN OPEN
AVOID EXPOSURE TO BEAM.
LASERSTRAHLUNG.
WENN ABDECKUNG GEÖFFNET
NICHT DEM STRAHL AUSSETZEN.
RAYONNEMENT LASER EN CAS
D'OUVERTURE
EXPOSITION DANGEREUSE AU FAISCEAU.
RAGGIO LASER IN CASO DI
APERTURA NON ESPORSI AL RAGGIO.
RADIACION LASER EN CASO DE APERTURA
EVITAR LA EXPOSICION A LOS RAYOS.
LASERSTRÅLNING OM ÖPPNAD
UNDVIK ATT UTSÄTTA DIG FÖR STRÅLEN.
FORSIGTIG
LASERUDSTRÅLING VED ÅBNING
UNDGÅ UDSÆTTELSE FOR STRÅLEN.
ADVARSEL
LASERSTRÅLER UTGÅR HVIS ÅPNES
UNGÅ EKSPONERING AV STRÅLEN.
EN60825-1:1996
Class
III
b Panel Label
Содержание FCR XG-1
Страница 6: ...vi 010 021 20 09 2001...
Страница 7: ...1 1 010 021 20 09 2001 1 Chapter 1 Introduction...
Страница 17: ...2 1 010 021 20 09 2001 2 Chapter 2 Operations...
Страница 37: ...3 1 010 021 20 09 2001 3 Chapter 3 When an Error Occurs...
Страница 39: ...A 1 010 021 20 09 2001 A Appendix A Specifications...
Страница 46: ...Specifications A 8 010 021 20 09 2001...
Страница 47: ...B 1 010 021 20 09 2001 B Appendix B IP Handling...
Страница 50: ...IP Handling B 4 010 021 20 09 2001...
Страница 51: ...C 1 010 021 20 09 2001 C Appendix C Film Annotation...
Страница 54: ...Film Annotation C 4 010 021 20 09 2001...
Страница 55: ...Y 1 010 021 20 09 2001 Y Appendix Y Exposed Anatomical Regions ApplicableMenus DisplayParameters...
Страница 91: ...Z 1 010 021 20 09 2001 Z Appendix Z Precautions for Exposure...
Страница 100: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...