Precautions for Exposure
Z-7
1
2
3
A
B
C
Y
Z
010-021-20 09.2001
2 Precautions for Exposure in the SEMI-AUTO MODE
The precautions are common to Semi I, II, III or III(**).
1
Center area of the IP
10
×
10cm (Semi I)
7
×
7cm (Semi II)
5
×
5cm (Semi III)
Position the portion you need to display often in each of the 5
×
5cm center areas of
the half-split IPs (both upper and lower halves and right and left halves) and the quar-
ter-split IP.
2
Never position anything other than the subject in the above mentioned areas. If any-
thing other than the subject is positioned in such areas, the image density will become
thinner.
In addition, do not position any metals or artificial bones in such areas. The image
density will become higher if such an object is positioned in such areas.
3
It is necessary to control the tube voltage as usual according to the subject thickness.
The following precautions should be observed for Semi IV.
( )
a
b
c
d
e
y
x
Area
a
b
c
d
e
Center Coordinate (x:y) cm
(0, 0)
(-5, 7)
(5, 7)
(-5, -7)
(5, -7)
Size (cm)
10
×
10
6
×
6
6
×
6
6
×
6
6
×
6
(1) Do not position transparent portions (area other than the subject) in the
abovementioned five areas.
(2) It is necessary to control the tube voltage as usual according to the subject thick-
ness.
The SEMI-AUTO MODE has been preset around the magnification menu. For details,
refer to Appendix Y Exposed Anatomical Regions / Applicable Menus / Display Pa-
rameter.
Содержание FCR XG-1
Страница 6: ...vi 010 021 20 09 2001...
Страница 7: ...1 1 010 021 20 09 2001 1 Chapter 1 Introduction...
Страница 17: ...2 1 010 021 20 09 2001 2 Chapter 2 Operations...
Страница 37: ...3 1 010 021 20 09 2001 3 Chapter 3 When an Error Occurs...
Страница 39: ...A 1 010 021 20 09 2001 A Appendix A Specifications...
Страница 46: ...Specifications A 8 010 021 20 09 2001...
Страница 47: ...B 1 010 021 20 09 2001 B Appendix B IP Handling...
Страница 50: ...IP Handling B 4 010 021 20 09 2001...
Страница 51: ...C 1 010 021 20 09 2001 C Appendix C Film Annotation...
Страница 54: ...Film Annotation C 4 010 021 20 09 2001...
Страница 55: ...Y 1 010 021 20 09 2001 Y Appendix Y Exposed Anatomical Regions ApplicableMenus DisplayParameters...
Страница 91: ...Z 1 010 021 20 09 2001 Z Appendix Z Precautions for Exposure...
Страница 100: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...