FujiFilm DR-ID 1200PB Скачать руководство пользователя страница 7

1-1

For Safe Operation

1

DR-ID 1200PB Operation Manual    

897N120972A

Chapter 1

  For Safe Operation

1.1  Safety

Before using the DR-ID 1200PB, read this section thoroughly to ensure that you use the product 
properly.

Electric Shock Warnings and Cautions

WARNING

The power supply to this device is AC100 to 240V.

To avoid electric shocks, users should always take the following precautions:

●  Do not open any covers when it is not necessary.

●  Install the equipment in a location where it will not be exposed to water.

●  Check that the equipment is securely earthed.

●  Check that all of the cables are completely and securely connected.

WARNING

Do not touch the patient’s body while touching the control cabinet and the image processing 

unit. Otherwise, the patient may receive an electric shock.

WARNING

Do not use a multiple tap connector or extension cable for powering the devices constituting 

the system. Otherwise, fire or electric shock may occur due to the electrical load exceeding the 

allowable limit.

WARNING

Observe the following precautions when using the cables. 

●  Do not touch the plug and connector with wet hands.  Otherwise, electric shock may result, 

causing death or severe injury. 

●  Hold the plug or connector when removing the cable.

  Pulling the cable or carrying by holding it may damage the cable, causing fire or electric shock. 

●  Do not damage or remodel the cable.   

Do not place a heavy object on the cable or lay it under the flat panel sensor. Do not step on, 

pull, forcibly bend, or bundle the cable. Otherwise, fire or electric shock may result. 

●  Do not use the flat panel sensor for the radiographic examination stand if its cable becomes 

overloaded. Otherwise, the cable may be damaged, causing fire or electric shock.

WARNING

Do not turn on the system with dew condensation on the flat panel sensor. Otherwise, fire or 

electric shock may result.

WARNING

Do not use the equipment in a location where metal particles could come into the equipment. 

This may cause an electric shock.

Содержание DR-ID 1200PB

Страница 1: ...Manual and other manuals for the related products before actually operating the power box After reading this manual store it nearby the power box so that you can see it whenever necessary For Safe Ope...

Страница 2: ...ii DR ID 1200PB Operation Manual 897N120972A...

Страница 3: ...not supplied by FUJIFILM Corporation 5 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from remodeling maintenance and repair using repair parts other than those specif...

Страница 4: ...iv DR ID 1200PB Operation Manual 897N120972A...

Страница 5: ...anel Sensor Connector 3 1 3 1 3 Changing the Direction of the Flat Panel Sensor Connector 3 2 3 1 4 Checking the Connection between the Flat Panel Sensor and the Connector 3 2 3 2 Starting Up and Shut...

Страница 6: ...vi DR ID 1200PB Operation Manual 897N120972A...

Страница 7: ...system Otherwise fire or electric shock may occur due to the electrical load exceeding the allowable limit WARNING Observe the following precautions when using the cables Do not touch the plug and co...

Страница 8: ...equipment are long be careful not to entangle the cables during use Also be careful not to trip over the cables Falls could result in injury CAUTIONS Follow the specified procedure when turning off t...

Страница 9: ...le produit En cas d endommagement de l quipement suite une chute ou un impact Installation Precautions CAUTIONS Do not install the equipment in a location with the following conditions Where the tempe...

Страница 10: ...Addition of devices Removal of devices Update of devices Upgrade of devices Warnings and Cautions on Network WARNING Make sure to use the optional parts and networks recommended by us Failure to use...

Страница 11: ...ns WARNING Confirm that the respiratory density of disinfectant including solvent is under legal regulation Certain disinfectants may damage health When using a disinfectant follow instructions suppli...

Страница 12: ...inspection while the equipment is used for a patient CAUTIONS The institution where the equipment is installed is responsible for its use and maintenance In addition this equipment should not be used...

Страница 13: ...DR ID 1200 The products in can be installed in patient environment Only one flat panel sensor can be connected to one power box in wired communication mode Up to three power boxes can be used for one...

Страница 14: ...product to a HUB and connect an additional LAN cable via the HUB 2 3 Locations of Labels and Signs Locations of labels and signs affixed to the DR ID 1200PB and the relevant safety signs are shown bel...

Страница 15: ...ase contact our official dealer or FUJIFILM Representative Year of manufacture Refer to Instruction Manual Booklet 2 3 3 Symboles de s curit et autres Les symboles de s curit suivants sont utilis s su...

Страница 16: ...2 4 System Configuration Product Overview 2 DR ID 1200PB Operation Manual 897N120972A...

Страница 17: ...sensors are used make sure that the READY lamp among the status lamps of the flat panel sensor to be used for an exposure is lit Connect Disconnect the connector straight to the flat panel sensor If c...

Страница 18: ...the connector of the SE cable to the flat pane sensor check the power ON OFF state and communication state with the status LEDs on flat panel sensor POWER Blue The power on off state of the flat pane...

Страница 19: ...hat the READY status lamp on the flat panel sensor is lit 3 2 3 Shutting Down the DR ID 1200PB 1 Disconnect the SE connector from the flat panel sensor and then disconnect the power cable of the power...

Страница 20: ...3 4 Basic Operation 3 DR ID 1200PB Operation Manual 897N120972A...

Страница 21: ...best conditions 4 1 1 Periodical Inspection Inspection Every Three Months Once every three months remove any dirt or dust accumulated in each part of the equipment using a vacuum cleaner or air duste...

Страница 22: ...4 2 Daily Inspection and Maintenance 4 DR ID 1200PB Operation Manual 897N120972A...

Страница 23: ...Humidity 15 RH 15 C 80 RH 30 C no dew condensation Atmospheric pressure 700hPa 1060hPa 2 Non operating Conditions Environmental conditions under which power can be supplied Temperature 5 C 35 C Humid...

Страница 24: ...below Specifications dimensions and weight are subject to change for improvement without prior notice A 2 1 DR ID 1200PB Width mm in Depth mm in Height mm in Weight kg lb Power box 259 10 2 205 8 1 2...

Страница 25: ...when this product is used in an examination car For the external view see O 1 1 Retaining Bracket for PB page O 1 O 1 1 Retaining bracket for PB The retaining bracket for PB is an optional item This...

Страница 26: ...O 2 Appendix O Use of Optional Items DR ID 1200PB Operation Manual 897N120972A...

Страница 27: ...peration Manual 897N120972A Maintenance and Inspection 1 Maintenance and Inspection Items Assigned to Specified Dealer For periodical inspection of the equipment and necessary arrangements consult our...

Страница 28: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...

Отзывы: