8. Specification
8.1 General specification
Type
NRF31021-
□
21YY , NRF31021-
□
22YY
Radiation type
Gamma (X)-ray ( 35keV to 6.0MeV)
Neutron
( 0.025eV to 15MeV)
Detector
Silicon semi-conductor
Dose range
0
µ
Sv to 9.999 Sv
Display
LCD
Calibration
±
10%
137
Cs, 1mSv (10mSv/h)
±
15%
252
Cf, 1mSv (10mSv/h)
Energy response
50keV to 1.5MeV
<
±
20% (Ref.
137
Cs)
1.5MeV to 6MeV
<
±
30% (Ref.
137
Cs)
250keV to 4.5MeV
<
±
50% (Ref.
252
Cf)
Dose accuracy
±
10%
( 0.1 mSv to 9.999 Sv) ( Gamma (X)-ray)
±
15%
( 0.15mSv to 9.999 Sv) (neutron)
Dose rate accuracy
±
20%
( 2 mSv/h to 9.999 Sv/h) (Gamma (X)-ray)
±
20%
( 150 mSv/h to 9.999 Sv/h) (neutron)
Linearity
±
10%
( 0.1 mSv/h to 9.999 Sv/h) (Gamma (X)-ray)
±
20%
( 0.5 mSv/h to 9.999 Sv/h) (neutron)
Angular of incidence
±
20%
( 0 deg. - 60 deg.)(
137
Cs, Ref. 0 deg.)
±
50%
( 0 deg. - 60 deg.)
(
241
Am or equivalent of 60keV X-ray, Ref. 0 deg.)
±
30%
( 0 deg. - 75 deg.) (
241
Am-Be, Ref. 0 deg.)
Ambient temperature
±
20%
(-10
℃
to +40
℃
) (Ref. 20
℃
)
Response time
5 sec or less
( 5 mSv/h to 9.999 Sv/h) (Gamma (X)-ray )
( 100 mSv/h to 9.999 Sv/h) (neutron)
Battery
Li-ion battery x1 (CR123A)
Battery life
2880 hours
(new battery, normal temperature, no alarms)
Operating temperature
-20
℃
to +50
℃
(90% or less, no condensation)
Dimensions
Approx. 60 X 78 X 33mm (including clip)
Weight
Approx. 125g
(including clip, rubber cover and battery)
Tests on the dosimeter comply with the IEC Standard IEC 61525 (1998), and 61526 (1998)
、
and JIS
Standard JIS Z 4312 (2003). Indications on the dosimeter may be interfered by certain factors such
as mobile phones, vibrations, and impacts.
If a hard impact operating on the dosimeter, it may have a crack to the dosimeter. In this case,
it may deteriorate the capabilities in a waterproof and a resistance of
radio wave.
Содержание NRF31
Страница 27: ...WTA528775 27 33 8 2 Energy response 8 2 1 Gamma X ray 8 2 2 Neutron Relative response X ray energy keV...
Страница 29: ...WTA528775 29 33 8 4 Angular of incidence 8 4 1 For 137 Cs on phantom 8 4 2 For 241 Am X ray 56 2keV on phantom...
Страница 30: ...WTA528775 30 33 8 4 3 For 137 Cs in Free air 8 4 4 For 241 Am Be on phantom...
Страница 31: ...WTA528775 31 33 8 5 Ambient temperature...