WTA528775 17/33
5. Battery Replacement
5.1 Battery replacement
Follow these steps to replace the battery :
(1) Insert the plug pin (*Option) to
turn the dosimeter off.
(2) Unscrew the battery cap using a coin.
(3) Replace the battery.
Insert a new battery in the compartment correctly.
(4) Close the battery cap using a coin till the
O-ring is disappeared.
An O-ring is consumables, please change the O-ring every 2 years or the when of battery
replacement (recommend). The information of the O-ring purchase, please contact our agency.
Attention
1. When replacing battery, make sure to turn off the dosimeter by inserting the
Plug pin.
2. During replacement, align the battery polarity correctly.
3. Use only CR123A battery in 3.0 V. *4
*4 The battery voltage is checked on electrodes of CR123A by using a ordinary voltmeter.
Becase of a load resistance in the dosimeter, in some cases, the measured voltage is higher
than the voltage during the operation.
Battery
(CR123A)
Plug pin
(Option)
Battery
cap
+
-
Содержание NRF31
Страница 27: ...WTA528775 27 33 8 2 Energy response 8 2 1 Gamma X ray 8 2 2 Neutron Relative response X ray energy keV...
Страница 29: ...WTA528775 29 33 8 4 Angular of incidence 8 4 1 For 137 Cs on phantom 8 4 2 For 241 Am X ray 56 2keV on phantom...
Страница 30: ...WTA528775 30 33 8 4 3 For 137 Cs in Free air 8 4 4 For 241 Am Be on phantom...
Страница 31: ...WTA528775 31 33 8 5 Ambient temperature...