WTA528775 16/33
Alarm activation
Alarm cause
Beep pattern
Overflow
Accumulate dose reaches a
10Sv or
Dose rate reaches a 10Sv/h.
・
Accumulate dose and dose rate (common)
The alarm begins with a 1 min-long beep, followed
by a 1 sec mute. After that, a 1 sec beep and a 1
sec mute are repeated. All beeps go on for
preset alarm duration. *3
LED is activated in 1 sec. interval.
*3 If the mode-switch is pressed in 3 second during preset alarm duration, the buzzer and LED
are inactivated. (This Alarm-Stop-Function can switch ON and OFF using a setting device.
Default setting is ON.)
Attention
If you press and hold the button until the audible/LCD alarm stops and then
dose rate alarm occurs, audible/LCD alarm for dose rate activates only when
the dose rate exceeds 10Sv/h.
As for accumulated dose alarm in this case, audible alarm does not activate
and only “AL” and “OVER” are blinking.
* One-second interval alarm sound does not activate in this case.
preset alarm duration
Alarm Type (blink)
LED
(
blink
)
●
Alarm Type (blink)
Содержание NRF31
Страница 27: ...WTA528775 27 33 8 2 Energy response 8 2 1 Gamma X ray 8 2 2 Neutron Relative response X ray energy keV...
Страница 29: ...WTA528775 29 33 8 4 Angular of incidence 8 4 1 For 137 Cs on phantom 8 4 2 For 241 Am X ray 56 2keV on phantom...
Страница 30: ...WTA528775 30 33 8 4 3 For 137 Cs in Free air 8 4 4 For 241 Am Be on phantom...
Страница 31: ...WTA528775 31 33 8 5 Ambient temperature...