33
5.4 (20) Status output check
Description
Perform check of status output for ON-OFF operation. Status output is an open collector. A
check is performed by connecting a voltmeter to terminals, TRout 1 and TRout 2 as shown
below.
Operation (example) Check of status channel 1.
Key operation
FUNC SYSTEM
or
or
, ENTER
or
ESC ESC ESC
Description
Select “Status check”.
Select “Channel 1”.
Select “ON or OFF”.
Press the key 3 times.
STATUS CHECK
Channel *
(Measurement display)
Display
STATUS CHECK
Channel 1
STATUS CHECK
ON
TR out 1
Volt meter
TR out 2
1
2
3
4
Volt meter
V
V
Signal receiving instrument
Note:
Status output changes depending upon “normal” or “Reverse” specified under the status
mode conditions.