Syntax for measurements with aC! or aM!
Command
Answer
aD0!
a<values><CR><LF>
a
– Sensor address
a
– Sensor address
D
– Command ‘Send Data’
<values>
– Requested data separated by
0
– Request for the data in bu
ff
er 0
resp. sign (‘+’ or ‘-’)
or
1
= bu
ff
er 1
!
– End of command
<CR><LF>
– End of answer
Example:
Command
Answer
0C!
000306<CR><LF>
0D0!
0+0.100+6.000+0.100<CR><LF>
0D1!
0+6.000+12.000+25.231<CR><LF>
Syntax for measurement with aCC! or aMC!
Command
Answer
aD0!
a<values><CRC><CR><LF>
a
– Sensor address
a
– Sensor address
D
– Command ‘Send Data’
<values>
– Requested data separated by
0 – Request for the data in bu
ff
er 0
resp sign (‘+’ or ‘-’)
or
1
= bu
ff
er
1
<CRC>
– 3-digit CRC checksum
!
– End of command
<CR><LF>
– End of answer
Additional Measurements
With the following commands additional information can be generated by the
RAINE-HYDRO
and requested
with
aD0!
.
The additional measurement commands
aMn!
and
aMCn!
have the same format as the
commands
aM!
and
aMC!
. The same applies to the commands
aCn!
and
aCCn!
which have the same format
as
aC!
and
aCC!
.
n
Function
Command
Answer
1
Generation of the variance (period 4 s )
aM1!
a0031<CR><LF>
aC1!
a00301<CR><LF>
with CRC
checksum
aMC1! a0031<CR><LF>
aCC1!
a00301<CR><LF>
2
Generation of the heating data
Measured variable
Unit
Bu
ff
er 0
Temperature interior
°C
Heating ON (1) / OFF (0)
Heating power total
%
aM2!
a0033<CR><LF>
aC2!
a00303<CR><LF>
with CRC
checksum
aMC2!
a0033<CR><LF>
aCC2!
a00303<CR><LF>
22
SDI-RAINE-HYDRO
USER MANUAL
700-SDI-RAINE-HYDRO Rev1 03 Mar 2021
Part# 21184