DigiTemp User Manual 

Chapter 2  SDI Commands 

 

page 11 

2.4.1

 

Start Measurement 

This command is used trigger a measurement on the addressed sensor. The sensor will not return data, instead 
the sensor will return the duration of the measurement as well as the number of data points returned by the 
measurement. The data is read using a subsequent Send Data command. Refer to Tables 2 and 3 for the 
specifics of the data returned from the sensor. 

 

String 

Note 

Cmd 

3M!

 or 

3MC!

 

start a measurement on sensor at address 3 (without or 

with CRC in data response, respectively) 

Resp 

30032 

3

 

device SDI address 

003

 

measurement delay (until data is ready; seconds) 

2

 

number of data points returned 

2.4.2

 

Start Concurrent Measurement 

The concurrent measurement command allows a sensor to take a measurement while other SDI sensors are 
also taking measurements. The Start Concurrent Measurement Command operates in the same manner and 
returns the same information as the Start Measurement command (refer to Table 2). 

 

String 

Note 

Cmd 

3C!

 or 

3CC!

 

start a concurrent measurement on sensor at address 3 

(without or with CRC in data response, respectively) 

Resp 

300302 

3

 

device SDI address 

003

 

measurement delay (until data is ready; seconds) 

02

 

number of data points returned 

2.4.3

 

Send Data 

This command requests the data generated by the preceding Measurement (

M

 or 

C

) command. An 

aD0!

 

command is always the first command sent to retrieve the data. If additional data needs to be read, then an 

aD1!

 

command is sent, then and 

aD2!

 

etc. etc., up to 

aD9!

. Since the DigiTemp returns only two values 

(temperature in degrees C and F), only an 

aD0!

 command is ever required with it. 

 

String 

Note 

Cmd 

3M! 

start a measurement on sensor at address 3 

Resp 

30032 

3

 

device SDI address 

003

 

measurement delay (until data is ready; seconds) 

2

 

number of data points returned 

Cmd 

3D0! 

request data from previous measurement command 

Resp 

3+11.4+52.5 

3

 

sensor SDI address 

+11.4

  temperature in degrees Celsius 

+52.5

  temperature in degrees Fahrenheit 

Cmd 

3MC! 

start a measurement on sensor at address 3, with CRC in 

data response 

Resp 

30032 

3

 

device SDI address 

003

 

measurement delay (until data is ready; seconds) 

2

 

number of data points returned 

Содержание DigiTemp

Страница 1: ...1 800 548 4264 www ftshydrology com 700 DigiTemp Rev 1 0 24 Nov 2011 DigiTemp SDI 12 Submersible Digital Temperature Sensor User Manual...

Страница 2: ...Contact information FTS 1065 Henry Eng Place Victoria B C V9B 6B2 CANADA www ftshydrology com Toll free 1 800 548 4264 Local 250 478 5561 Technical support support ftshydrology com...

Страница 3: ...Revision History Date Person Description 24 Nov 2011 RG First released version 28 Nov 2011 RG Late corrections...

Страница 4: ...3 Connection 7 1 4 Operation 7 1 5 Configuration 7 1 6 Calibration and maintenance 7 Chapter 2 SDI Commands 8 2 1 Notation for SDI commands 9 2 2 Commands recognized by DigiTemp 9 2 3 General SDI com...

Страница 5: ...Chapter 1 Operation Chapter contents 1 1 General description 6 1 2 Installation 6 1 3 Connection 7 1 4 Operation 7 1 5 Configuration 7 1 6 Calibration and maintenance 7...

Страница 6: ...these steps 1 Place the sensor in a location where it is continuously immersed in a representative sample of the water to be monitored usually a well mixed locale 2 The sensor may be mounted on an ob...

Страница 7: ...s powered by the 12 V DC power supply on the SDI 12 bus For details on the SDI commands used by the sensor see Chapter 2 SDI Commands 1 5 Configuration The DigiTemp is shipped with default address 0 u...

Страница 8: ...2 3 General SDI commands 9 2 3 1 Address Query 2 3 2 Acknowledge Active 2 3 3 Change Address 2 3 4 Send Identification 2 4 SDI Data commands 10 2 4 1 Start Measurement 2 4 2 Start Concurrent Measureme...

Страница 9: ...Commands recognized by DigiTemp The table below provides a quick reference to the commands recognized by the DigiTemp For details on these commands see the following sections Command name Command code...

Страница 10: ...address String Note Cmd 0 request the device at address 0 to confirm it is active Resp 0 a device is present at address 0 2 3 3 Change Address This command changes a sensor s SDI address String Note C...

Страница 11: ...Table 2 String Note Cmd 3C or 3CC start a concurrent measurement on sensor at address 3 without or with CRC in data response respectively Resp 300302 3 device SDI address 003 measurement delay until d...

Страница 12: ...ata is returned Cmd 1RC perform a continuous measurement read on sensor at address 3 with CRC code in data response Resp 1MSA 1 sensor SDI address MSA CRC code DigiTemp is not a continuous measurement...

Страница 13: ...Chapter 3 Technical specifications...

Страница 14: ...may be less accurate Responsiveness Time constant 0 57 min reaches 95 of final value in 1 7 min 99 in 2 9min Measurement time 3 s Power supply voltage 12 V DC nominal range 9 6 V 18 6 V Standby curre...

Отзывы: