FSN2006 - 4/2019
English - 15
5. Recommended separation distances between portable and mobile RF
communications equipment and the medical LCD monitor
The medical LCD monitor is intended for use in the electromagnetic environment in which the RF distur-
bances are controlled. The user of the device can help prevent electromagnetic interference by maintai-
ning a minimum distance between portable and mobile RF communications equipment (transmitters) and
the device – as a function of the output power of the communication device, as shown below.
Nominal power of
transmitter [W]
Separation distanced [m] according to frequency of transmitter
150kHz to< 80 MHz
d = 1.2 P
80 MHz to< 800 MHz
d = 1.2 P
800 MHz to 2.5 GHz
d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation dis-
tance
d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,
where
P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.