5.
Installation
5.1
Delivery - Unpacking
Use a pallet truck to move the freezer on its pallet. The freezer must be supported during unpacking to avoid any risk
of toppling. The unit may be placed on the floor and manoeuvred on casters. Once the unit has been placed in the
desired location, remove the protective plastic film and chocks.
Froilabo freezers are delivered on a pallet with an unloading ramp. This means that no special equipment is required
to unload them from the pallet. Please refer to the handling and unpacking sheet affixed to the unit.
Do not use sharp objects as they might damage the paintwork.
Note: After unpacking, inspect (if possible with the carrier present) the condition of the unit and its accessories. If any
transport damage is observed, state the details on the delivery note and inform Froilabo without delay.
After receipt, please check the items delivered:
BM 3E – 340 litres
BM 3E – 515 litres
BM 3E – 690 litres
1 non-detachable electrical power lead
1 filter cassette
1 set of two keys
1 pressure relief valve + filters
1 manual
1 shelf
2 shelves
3 shelves
2 brackets
4 brackets
6 brackets
5.2
Environment conditions
The freezer is designed for use under the following ambient conditions (in accordance with EN 61010-1):
Indoor use
Maximum altitude: 2000 m
Ambient temperature range between 5°C and 40°C
Maximum relative humidity of 80% for temperatures up to 22°C
Supply voltage ripple < ±10 % of rated voltage.
Supply network voltage surges: category II (Standard IEC 60364-4-44)
Maximum level of room contamination: 2
BM 3E freezers are Class A devices within the meaning of the EMC Directive.
5.3
Electrical environment
A 230 V / 50 Hz or 120 V / 60 Hz single-phase electrical power supply is required, as specified on the manufacturer’s
plate at the rear of the unit.
The freezer must be powered via a safety mechanism, such as a correctly rated differential circuit breaker, which
automatically cuts off the power supply if an insulation fault occurs.
To enable the unit to be electrically cut off, the circuit breaker must be readily identifiable and within reach of the
operator.
5.4
Location
To minimise power consumption and achieve the specified performance, place the freezer in a ventilated location,
away from sources of heat (radiators, heaters, etc.) and avoid locations exposed to sunlight. Place it on a flat surface.
For correct operation, it is essential that the freezer be placed level. Use rigid chocks if necessary.
Ensure that there are no obstacles (walls, equipment) that impede the ventilation of the freezer (air inlet and outlet).
Lower the two front legs, turning by an additional quarter-turn after they touch the ground.