8
Maintenance
DANGER!
Mains voltage
– Before beginning with maintenance work, unplug
the mains plug and protect the device against being
unintentionally switched back on again!
– Indicate maintenance work with warning signs.
– Maintenance work may only be performed by spe-
cialised personnel.
– Put safety equipment back into operation after
maintenance or repair work.
We recommend keeping a safety logbook
Ä
Chapter 13
„Safety logbook“ on page 52, where all work (mainte‐
nance, repairs......) performed on the device is entered.
The most important element of maintenance is regular
cleaning:
Functional part
Task
Test
Maintenance interval
Safety lock
Hood lock
Is the closed hood locked in
place when the main
switch is off?
If this test is failed, do not
continue to work until the
fault has been rectified.
Before each use
LID LOCK ACTIVATED light
on the control panel
Shows that the hood is
closed
When the hood is closed
the LID LOCK ACTIVATED
light should light up green
Once a month
Rotating bearings
Permanent lubrication
Bearing clearance
Every 2,000 h or annually
Drive motor
Permanent lubrication
Bearing clearance
Every 4,000 h or annually
V-belt
Motor to planetary disc
Check the voltage; remove
the housing; the belt must
not slacken by more than
approx. 10 mm when
pushed with your thumb.
Once a year
Fan
Grinding chamber cooling
and electronics
Proper function, clean
when soiled
Twice a year
Maintenance
- 43 -