3
Technical data
3.1 Dimensions
Models I + II: 72 x 41 x 83 (height x width x depth)
3.2 Weight
Model I:
177 kg net
202 kg gross
Model II:
205 kg net
230 kg gross
3.3 Operating noise
Emissions value of workplace according to DIN EN ISO 3746:2005 L
Pa
= 93
dB(A).
3.4 Voltage, current consumption, power consumption
The device can be operated in the following voltage ranges:
Model I:
n
Single-phase alternating current 115 V +/- 10 %, 21 A, 1.9 kW
n
Single-phase alternating current 230 V +/- 10 %, 9 A, 1.57 kW
n
Three-phase alternating current 200 V +/- 10 %, 5.3 A, 1.1 kW
n
Three-phase alternating current 400 V +/- 10 %, 2.62 A, 1.45 kW
Model II:
n
Three-phase alternating current 230 V +/- 10 %, 9 A, 2.78 kW
n
Three-phase alternating current 400 V +/- 10 %, 4.95 A, 2.2 kW
n
Three-phase alternating current 500 V +/- 10 %, 5.15 A, 3 kW
n
Single-phase alternating current 230 V +/- 10 %, 13.5 A, 2.36 kW
Transient overvoltages in accordance with overvoltage category II are
permitted. (See also
Chapter 4.6 ‘Electrical connection’ on page 20)
3.5 Electrical fuses
The electrical fuse is a motor protection switch, which is integrated in the
main switch.
Technical data
- 15 -
Содержание PULVERISETTE 1 classic line
Страница 51: ...13 Safety logbook Date Maintenance Repair Name Signature Safety logbook 51 ...
Страница 52: ...Date Maintenance Repair Name Signature Safety logbook 52 ...
Страница 55: ......