6.3 Setting the amplitude
Selecting the amplitude:
n
The vertical oscillation amplitude can be set in 0.1 mm steps using
the + / - keys. Values between 0.1 and 3 mm can be set with the
PULVERISETTE 0.
NOTICE!
Grinding at an amplitude higher than 2 mm can cause
damage to the device and the grinding set.
6.3.1 Displaying the amplitude
The amplitude must be checked on the amplitude plate on the clamping
lid. The lines that appear to touch one another display the set amplitude.
The principle is based on the persistence of vision of the human eye.
Example for 2 mm amplitude:
Using the device
- 25 -