7
Accessories
7.1 Sound absorption hood
In order to reduce the operating noise of the micro mill a sound absorp-
tion hood (Art. no. 00.0130.17) is available.
7.2 Conversion to Vibratory Sieve Shaker ANALYSETTE 3 SPARTAN
The PULVERISETTE 0 can also be used as an ANALYSETTE 3
SPARTAN by using a sieve pan, test sieves and clamping lid.
The ANALYSETTE 3 SPARTAN is used to sieve dry bulk solids as well as
particle collectives in suspensions. Depending on feed quantity and grain
size, the test sieves and sieve pans can be clamped with different diame-
ters and heights.
7.2.1 Conducting a sieve analysis
7.2.1.1
Sieve mesh widths that can be used
n
Dry sieving
100 µm to 25 mm (without sieving aids); 32 µm to 25 mm (with
sieving aids)
n
Wet sieving
20 µm or greater
n
Micro sieving (PRO version only)
5 µm to 100 µm
7.2.1.2
Fitting and clamping the sieves
On the vibratory plate, it is possible to fit up to
‐ 10 sieves with a height of 50 mm (or 2") or
‐ 16 sieves with a height of 25 mm (or 1")
between the sieve pan (collecting vessel) and clamping lid.
The combination of sieves and sieve pan is called a sieve set.
Make sure not to clamp more than 6 kg on the ANALYSETTE
3 (sieve set + sieving stock).
Accessories
- 26 -
Содержание PULVERISETTE 0
Страница 6: ...13 Exclusion of liability 46 14 Safety logbook 48 15 Index 50 Table of contents 6 ...
Страница 15: ...2 8 3 Overload protection The mains fuse provides overload protection Safety information and use 15 ...
Страница 40: ...NOTICE The plexiglas lid must not be cleaned with alcohol or organic solvents Cleaning 40 ...
Страница 48: ...14 Safety logbook Date Maintenance Repair Name Signature Safety logbook 48 ...
Страница 49: ...Date Maintenance Repair Name Signature Safety logbook 49 ...
Страница 51: ......