M-7.3-16 Rev. 05 09/03/15                 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER 

3

 

 

                                                                            Use and maintenance manual 

Index 

 

 

 

 

 

 

 



Products applicable to this manual                                       5 



Table of capacity and dimensions                                         6 



Technical data                                                                         6 

 



Standard and optional fitting                                                  6 

 

 

 

Standards and general warnings                                           7 

 

1.1 

Certification 

1.2 

Testing and warranty  

1.3 

Aim, contents and intended user of this manual  

1.4  

Product description  

1.5 

General safety regulations 

1.6 

Client’s responsibilities 

1.7 

Client  service requests 

1.8 

Ordering spare parts 

1.9 

Aim and intended use of device 

1.10 

Unsuitable conditions for use 

1.11 

Materials and refrigerants 

 

Installation                                                                             10 

 

2.1 

Transport and handling

 

2.2   

Positioning 

2.3      Wiring and electrical hook-up 
2.4      Set-up operations 
2.5   

Re-installation 

2.6     Scrapping and disposal 

 

 

 

 

 

 

 

 

 

 

 

 

Operation  

 

 

 

 

         

               14     

3.1     Safety and accident prevention 
3.2     Safety data plates and guards 
3.3     Operating limits 
3.4 

Storage conditions 

User

’s instructions   

 

 

 

                        15 

4.1      Start up of the appliance 
4.2      Functions 
4.3      Control panel 
4.4      The menu tree 
4.5      Alarm codes 
4.6      Download data procedure 
4.7      Procedure to disable data logger 
4.8      Procedure to calibrate internal temperature probe S1 
4.9      Update of firmware release 
4.10 

GSM module 

4.11 

Dry contact for remote alarm 

4.12    Back-up battery 
4.13    Temperature chart recorder 
4.14    Drawer 
4.15    Divider for drawer 
 

Содержание AF140E

Страница 1: ...M 7 3 16 Rev 05 09 03 15 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER 1 Use and maintenance manual USER PASSWORD 555 ...

Страница 2: ... out in conformity with the instructions the device is used in accordance with this manual COPYRIGHT NOTES All rights reserved No part of this document is to be reproduced stored or transmitted in any form without the prior written permission of Fri med s r l Despite every care being adopted during preparation of the document Fri med s r l shall not assume any responsibility for atrocities or negl...

Страница 3: ...ions for use 1 11 Materials and refrigerants 2 Installation 10 2 1 Transport and handling 2 2 Positioning 2 3 Wiring and electrical hook up 2 4 Set up operations 2 5 Re installation 2 6 Scrapping and disposal 3 Operation 14 3 1 Safety and accident prevention 3 2 Safety data plates and guards 3 3 Operating limits 3 4 Storage conditions 4 User s instructions 15 4 1 Start up of the appliance 4 2 Func...

Страница 4: ...rator 5 2 Cleaning the interior and exterior of the appliance 5 3 Cleaning the condenser 5 4 Precautionary measures for prolonged disuse 6 Special maintenance and repairs 30 7 Service Troubleshooting 31 8 Symbols used 32 9 Annexes 33 9 1 CE declaration of conformity 9 2 Technical documentation 9 2 1 Electric wiring diagram 9 2 2 Safety test report ...

Страница 5: ...th June 1993 concerning medical devices and following amendments Blood Bank Refrigerators SB FS PN AF MODELS Operating temperature 2 C to 6 C Factory pre set to 4 C Models SB10E FS15E FS20E FS25E FS30E PN45E PN52E AF70E AF140E Plasma Freezer CL60B PF MODEL Operating temperature 10 C to 42 C Factory pre set to 40 C Model CL60B PF All relevant data referring to FRI MED products can be found on the d...

Страница 6: ... SB10E 100 230 50 Hz 245 R404a 0 150 kg FS15E 150 230 50 Hz 245 R404a 0 200 kg FS20E 200 230 50 Hz 245 R404a 0 250 kg FS25E 250 230 50 Hz 360 R404a 0 270 kg FS30E 300 230 50 Hz 360 R404a 0 290 kg PN45E 450 230 50 Hz 295 R404a 0 250 kg PN52E 520 230 50 Hz 295 R404a 0 250 kg AF70E 700 230 50 Hz 415 R404a 0 250 kg AF140E 1400 230 50 Hz 510 R404a 0 340 kg CL60B PF 600 230 50 Hz 580 R404a 0 600 kg Powe...

Страница 7: ... This manual has been prepared with the objective to supply all instructions required for the correct use of the appliance and to keep it in optimal condition It also contains important user safety information The following professional roles are explained in order to define the responsibilities of each involved parties Installer a qualified technician who installs the appliance in accordance with...

Страница 8: ...the unit of any of these situations Primary general safety regulations Do not touch the unit with wet hands and or feet Do not use the appliance with bare feet Do not insert screwdrivers or other pointed objects between guards or moving parts of the appliance Do not pull the power cord to disconnect the appliance from the electrical mains Make sure that the appliance is not used by unsuitably qual...

Страница 9: ...OT COME OUT OF THE SHELF PERIMETER All uses except authorized used of the appliance shall be construed as improper use for which the manufacturer declines all responsibility 1 9 UNSUITABLE CONDITIONS FOR USE The appliance must not be used exposed to bad weather with adapters multiple sockets or extension leads in an explosive atmosphere or in an atmosphere with a fire risk or closed to heating sou...

Страница 10: ... from sources of heat Avoid direct sunlight Remove packing material Remove accessories from inside the unit Wood basement removal by means of a hammer until the feet block hinder ledge tilt the cabinet to one side and loosen the two thread forming screws by means of a socket wrench drag the cabinet from the back side holding the basement still until the four feet have gone out from the containing ...

Страница 11: ...MINARY CHECKS Check the condition of the power cord no cut or chaffing Check that the feed door hinges and shelf support are stable Check the door seals and drawers are not damaged broken or scratched and that the door closes and is sealed properly Make sure pipelines unions are in perfect condition INDICATIONS FOR OPTIMAL DUTY Do not block the motor compartment air vents Do not lay objects on the...

Страница 12: ...inated by your Local National Health Department or as specified by the law in force Scrapping and disposal of the appliance must be carried out in full observance of established legislation in your country In particular the apparatus may contain the following materials Iron Copper Aluminium Non biodegradable plastics Fibre glass for printed circuits Ferrite Batteries CFC free refrigeration gas Ele...

Страница 13: ...esigned and built to reduce risks related to the emission of airborne noise to the minimum level 3 2 SAFETY DATAPLATES AND GUARDS It is strictly forbidden To tamper or remove the evaporator cover that protects the user from the risk of cutting on the heat exchanger fins To remove the data plate fixed to the inside edge of the motor housing showing technical specification and earth connection warni...

Страница 14: ...word is correct after entering the last digit the appliance switches off 4 2 FUNCTIONS ANTIPANIC FUNCTION In all menu screens after 120 seconds the key button has been pressed the controller returns to the main menu without saving any unconfirmed changes LCD FUNCTION ENERGY SAVING One minute after pressing any key the backlight of the screen switches off automatically stand by status Lighting will...

Страница 15: ...DECODING F1 set point key temperature setting arrow F2 display of alarms memory arrow F3 light key arrow F4 menu ESC TEMPERATURE SETTING Pressing key F1 SET set point is shown on the display Pressing any of the arrow keys a message enter password XXX appears If password inserted is correct pre set data appears to be modified By using arrow keys the requested value can be set press ENTER to confirm...

Страница 16: ...y ESC you go back to the main menu 1 ALARMS MEMORY List of latest 20 recorded alarms with data time and type of alarm 2 INFO ENTRY Following data can t be modified Manual defrost Serial number Delay door Alarm delay open door alarm how many minutes before alarm starts High temp alarm delta T value difference with set point temperature and delay how many minutes before alarm starts Low temp alarm d...

Страница 17: ...vels 1 User end user 2 Assistance technician 3 Factory factory setting Enter password If password is correct parameters changes are enabled according to password level If password entered is not correct message wrong password will appear on display for five seconds and then it will automatically go back to previous menu To modify password data enter standard user password 555 then select key PASSW...

Страница 18: ...k Damaged clock 15 AM1 General alarm Condenser cleaning 16 AM2 General alarm check maintenance on appliance is required 17 AM3 Security Full Memory download records When the appliance is on alarms and data of probes are recorded Each single record is related to date and time It s possible to record a maximum of 46000 entries which corresponds to about one month recording Then you have to download ...

Страница 19: ...ata parameters setting 3 Data load enter new parameters F1 2 3 corresponds to arrows as indicated in the picture 1 Download records To download memory records press key ENTER and confirm it the message DOWNLOAD RECORDS is flashing to indicate operation in progress Once procedure is finished the message DOWNLOAD RECORDS will be replaced by message OFF KEY By removing the USB pen the display shows t...

Страница 20: ...R 5 ENTER 5 ENTER In this way you enter the MENU Scroll up to point 0 USER PARAMETERS ENTER Then scroll up to point 5 CALIBR PROBES1 ENTER At this point insert the new value and press ENTER to confirm Exit programming by means of ESC key to return to initial display screen Example on how to choose the new parameter to enter if the temperature showing on display is 4 C while inside the appliance wi...

Страница 21: ...itch on the appliance again J In some cases the word PARAMETER ALARM can show depending on FW release In order to reset the alarm press the key ALL third from left side below display MUTE key Repeat the operation until the alarm disappears At this point double check that new Firmware release number has been installed 4 10 GSM MODULE optional When an alarm is activated a message is forwarded to pre...

Страница 22: ...m condition with a remote alarm installation on your site The connection for remote alarm is a clean switch contact The contact exchanges under all alarm conditions Note connections have to be made as indicated below and above all should be tested with equipment in operation 4 12 BACK UP BATTERY The battery back up allows to monitor the temperature of the appliance and any faults and alarms in cas...

Страница 23: ...r with the key Fig 6 2 Push the spring of the chart holder sideways and lift it Fig 7 Fig 8 3 Remove the chart Fig 9 Fig 10 4 Insert the new chart 5 Lower the spring of the diagram holder and push it sideways Fig 7 Fig 8 Fig 9 Fig 10 HOW TO REPLACE THE FIBRE TIPPED PEN the fibre tipped pin is a consumable which has an average duration of about one year 1 Open the cover with the key Fig 6 2 With on...

Страница 24: ...battery 1 5 V LR03 and put another new one taking care of polarity Fig 11 5 Lower the pen 6 Close the cover with key Fig 11 PEN ARM CALIBRATION 1 We suggest to check recorder s calibration every 6 months 2 If there is a difference greater than 2 between the digital display and chart recorder proceed as follows 3 Take gently the arm B of the pen 4 With a small screwdriver calibrate the regulation s...

Страница 25: ...14 DRAWER All blood bank refrigerators are standard fitted with drawers according to the scheme at page 6 Each drawer has a maximum capacity of 20kg In case you need to remove the drawer for cleaning follow the steps outlined below in the photos Beware open only one drawer at a time fig 13 fig 14 fig 15 fig 16 fig 17 fig 18 ...

Страница 26: ...uipped with Plexiglas dividers optional accessory that allow the creation of longitudinal compartments inside the drawer In case you bought the kit of dividers for drawer you can proceed with the installation of the same following the steps outlined below in the photos The same procedure has to be followed in case of need to remove the partitions for thorough cleaning fig 19 fig 20 ...

Страница 27: ...he appliance is thoroughly cleaned in our factory before delivery We recommend however that you clean the interior of the appliance before use Before cleaning the appliance make sure the power cord is disconnected 5 2 CLEANING THE INTERIOR AND EXTERIOR OF THE APPLIANCE To this end we indicate cleaning products to be used for exterior and interior of the appliance Water and neutral detergent and no...

Страница 28: ...1 and remove any dust or fluff from the heat exchanger fins After cleaning start the appliance During this operation use the following personal safety measures safety glasses respirator mask gloves Fig 21 Compressor at base To access the condenser at base unscrew the 2 screws by means of a screwdriver fig 22 then pull with both hands the panel at the front bottom of the appliance fig 23 Fig 22 Fig...

Страница 29: ...e refrigerator should remain unused for more than 15 days please proceed as follows Switch off the appliance and disconnect the power cord Clean the interior of the cabinet shelves trays runners and supports paying special attention to critical areas such as articulations and magnetic sealing strips Leave doors slightly open to prevent accumulation of residual humidity ...

Страница 30: ...e manufacturer The manufacturer declines all liability in the case of work performed by the user or unauthorised persons or if non original spare parts are fitted to the appliance In case of special maintenance of blood bank refrigerator internal compartment the head of the department that deals with storage of blood is required to make a transfer of the entire contents of the refrigerator in anot...

Страница 31: ...r Set new temperature Wait for the end of defrost cycle switch off and then on again after few minutes Contact service department Contact service department Refrigeration unit runs constantly but does not reach set temperature Room temperature too high Condenser dirty Refrigerant needs recharging Fan of condenser does not work Door gasket is damaged Evaporator is coated with ice Defrost valve is o...

Страница 32: ...AND PLASMA FREEZER 32 Use and maintenance manual 8 SYMBOLS USED Remove the tension before removing the protection Warning symbol Earth symbol It is necessary to consult the Instruction Booklet before carrying out the operation Clean condenser regularly ...

Страница 33: ...OD BANK REFRIGERATORS AND PLASMA FREEZER 33 Use and maintenance manual 9 ANNEXES The following documentation is enclosed 8 1 CE DECLARATION OF CONFORMITY 8 2 TECHNICAL DOCUMENTATION 8 2 1 Electric diagram 8 2 2 Safety test report ...

Страница 34: ...ev 05 09 03 15 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER 34 Use and maintenance manual FRI MED S R L VIA RIVA 26 19 14021 BUTTIGLIERA D ASTI AT ITALY Tel 39 011 9921161 Fax 39 011 9921160 e mail info frimed it ...

Отзывы: