4 USER
’S INSTRUCTIONS
M-7.3-16 Rev. 05 09/03/15 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER
19
Use and maintenance manual
In case GSM Module and back-up battery are connected, alarm message is forwarded
automatically to memorized mobile numbers. When alarm is no more present, a notification
message of “standard” working will be forwarded.
IMPORTANT NOTE - whenever an alarm is shown on display, you need to press any key so
that the display is on and then select keys ALARM and MUTE to clear it.
In this way the alarm is removed from display, although it will remain in memory of Alarm
Menu.
IMPORTANT NOTE - Appliances without back-up battery
In case of power failure, when power returns, alarm AL00 (power failure) appears on
display, indicating that the unit has been without power and showing the time when power
failed.
To clear the alarm from display, press any key so that the display is on and then select keys
ALARM and MUTE. In this way the alarm is removed from display (it will remain in memory
of Alarm Menu).
4.6 DOWNLOAD PROCEDURE
Insert the USB key into the hole, the symbol «
» appears on the
display with related messages :
1 - Download records ( = working temperature and alarms recording)
2 - Download data (= parameters setting)
3 - Data load (= enter new parameters)
(F1-2-3 corresponds to « arrows » - as indicated in the picture)
1
–
Download records
. To download memory records, press key « ENTER » and confirm it, the
message « DOWNLOAD RECORDS » is flashing to indicate operation in progress.
Once procedure is finished, the message « DOWNLOAD RECORDS» will be replaced by
message « OFF KEY» . By removing the USB pen, the display shows the main menu screen.
In order to read the file in a correct way on your PC, please open it by using the Note Pad
program
. At the end of download record procedure, a file is generated containing information
according to the following format:
M;29-05-12;09:53;4,0;26,5;
Ex :
R
(= initial recording/appliance switch on)
Ex:
M
(=standard functioning)
; 29-05-12
(=date);
09:53
(=hour);
4,0
(=internal temperature);
26,5
(=condenser temperature)
In case an alarm is recorded, the file will appear as follows:
A;29-05-12;09:53;5;
Ex:
A (
=alarm);
29-05-12
(=date);
09:53
(=hour);
5
(=type of alarm)
To identify type of alarm, please refer to section 4.5
2 - Download data (parameters for service)
-
If you want to download parameters, that is the
setting of the refrigerator, press ENTER on Download; you will be requested to digit the Password.
Once the password is entered the data will be downloaded. When you read
“FINISHED” on
display, please remove the USB pen and the display will automatically return to the initial screen.
5,5 °C
F1
1
F2
F3
F4
Date
ESC