6
Operation
What if the dwell time is shorter
than expected?
In Treatment Based mode, your cycler will adjust the dwell time based on the total time programmed,
the number of cycles, as well as the time it takes for each drain and fill. If a drain or fill cycle takes
longer than estimated by your cycler, the dwell time will be shortened to make sure you end
treatment on time.
In Cycle Based mode, your dwell time will not be shortened.