Liberty Cycler Therapy Settings
What are the therapy ranges for the Liberty cycler?
•
Total therapy time
– Programmable up to 47 hours and 59 minutes
•
Fill volume range
– 500 mL to 4000 mL (pause and last fill volumes may be 0)
•
Dwell time
– As much as 9 hours and 59 minutes
•
Bag sizes allowed
– 2,000 mL, 3,000 mL or 5,000 mL
(the heater bag must hold the first fill volume plus 1000 mL)
•
Number of exchanges
– One to 99
•
Number of pauses
– You may or may not be prescribed pauses. If you are,
your PD Nurse will instruct you on the number of pauses to enter.
•
Maximum total volume
– Each treatment can deliver up to 25,000 mL
•
Drain Time
– Lowest programmable drain time in minutes = fill volume divided by 100;
maximum drain time = 59 minutes
•
patient weight
– The dry weight may be set between 44 lbs (20 kgs) up to 880 lbs (400 kgs)
•
Daytime manual exchange
– 0 to 4,000 mL