6. Hydraulic leakage (treatment not possible)
70ml of fluid has inte nally leaked in the hydraulics.
Discontinue treatment and contact Technicians for assistance.
7. Micro bubble alarm
Follow on screen instructions, use operating manual if necessary.
8. Air handling procedure
Disconnect patient and re-circulate blood lines following the
on-screen instructions.
9. Venous pressure too high – reduce pressure manually
Open right hand door, attach 20ml syringe to venous bubble catcher,
unclamp and release pressure, alarm should clear automatically.
0462-FMC-5008-TS-Brief-OpGuide.indd 13
09/09/2016 11:43