1.
When the patient access is prepared and the full prime
volume has been reached, stop the prime by pressing the
blood pump button on the screen.
2.
Message ‘choices’ – ‘continue’ (rinse) or ‘exit’ (rinse) are
displayed, press ‘exit’ (rinse).
3.
Follow on screen instructions – connect arterial and venous
lines to patient access. Connect the SafeLine (sub tubing)
to the post or pre-dilution connector
F
/
G
.
4.
Ensure the Rinse port
E
is closed, close the EBM doors,
open clamps and confirm blood pump start
5.
Dialysis commences when blood is sensed by the
optical detector
H
below venous bubble catcher,
message ‘start dialysis treatment’ – press OK.
6.
Increase blood flow to the p escribed speed
7
.
7.
UF starts automatically if data is already set.
8.
Heparin starts automatically.
9.
Bolus may be given manually from heparin menu.
0462-FMC-5008-TS-Brief-OpGuide.indd 7
09/09/2016 11:43
Содержание 5008
Страница 1: ...The 5008 Therapy System Brief Operating Guide 62 FMC 5008 TS Brief OpGuide indd 1 09 09 2016 11 4 ...
Страница 16: ...Notes 62 FMC 5008 TS Brief OpGuide indd 16 09 09 2016 11 4 ...
Страница 17: ...Notes 62 FMC 5008 TS Brief OpGuide indd 17 09 09 2016 11 4 ...
Страница 18: ...Notes 62 FMC 5008 TS Brief OpGuide indd 18 09 09 2016 11 4 ...
Страница 19: ...Notes 62 FMC 5008 TS Brief OpGuide indd 19 09 09 2016 11 4 ...
Страница 20: ...Notes 62 FMC 5008 TS Brief OpGuide indd 20 09 09 2016 11 4 ...
Страница 21: ...Notes 62 FMC 5008 TS Brief OpGuide indd 21 09 09 2016 11 4 ...