5.6 Product Manifold Replacement
1.
Remove the PSA system per Section 5.4.
2.
Disconnect the tubing assembly from the two barbs on the product
manifold, mixing tank, and one end of the purge tube.
(Figure 5.20 a, b, c)
Figure 5.20a
Figure 5.20b
Figure 5.20c
3.
Remove the tie wraps that bundle the solenoid value wire together.
4.
Loosen and remove the center bolt from the PSA assembly using a 5/32”
Allen wrench. (