Instructions for Use (R-720-175, Rev. D)  

 

 

 

 

 

 

 

English

 

 

Due  to  the  posterior  placement  of  the  Kinterra’s  ankle  pivot  point,  changes  to  the 

plantarflexion  and  dorsiflexion  resistance  settings  will  be  most  noticeable  to  the  patient 
when  they  are  walking  on  slopes,  not level  ground.    Adjustments  of  the  resistances  are 
best done on a gradual slope as a final step in dynamic alignment.    
 
Start the dynamic alignment  by setting the heel stiffness first, to optimize the heel strike 
to mid-stance motion before proceeding on to adjust the dorsiflexion adjustment.  
 
Adjust  plantarflexion  resistance  to  dampen  heel  strike.    If  a  foot  slap  is  noticed  or  heel 

strike  it  too  abrupt,  increase  the  “PF”  resistan

ce.    If  the  heel  is  too  firm  or  the  knee  is 

buckling at heel strike, 

decrease “PF” resistance. 

 

 

Dorsiflexion 

“DF” resistance is adjusted to control the advancement of the amputee over 

the  foot.    If  the  patient  feels 

they are  walking “downhill”,

  increase  DF  resistance.    If  the 

effort to advance over the foot is challenging for the user, 

decrease “DF” resistance.

 

      

 

                                                                               INCREASE              DECREASE  

                                                                               

Resistance

              

Resistance   

                                                     

 

 
 
 
 
 
 

                                                                                 Smaller dot = lower resistance 
 

 

 

 

 

                     Larger dot = higher resistance 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание Kinterra

Страница 1: ...impact transtibial and transfemoral K3 level ambulators It is recommended for unilateral use Caution should be used when fitting bilateral patients due to the added movement of the ankle Kinterra has a weight limit of 275 lbs 125 kg Assembly The Kinterra foot ankle system is pre assembled consisting of a foot module ankle unit Spectra sock and foot shell After dynamic alignment torque pyramid adju...

Страница 2: ...exion is properly accommodated that the system still has 2 dorsiflexion and 10 plantarflexion Dynamic Alignment Once a stable static alignment is achieved begin walking the patient in the parallel bars to optimize alignment and function on level ground Using your patient s feedback adjust the hydraulic valves carefully to balance the comfort provided by the ankle damper and energy return provided ...

Страница 3: ...ke to mid stance motion before proceeding on to adjust the dorsiflexion adjustment Adjust plantarflexion resistance to dampen heel strike If a foot slap is noticed or heel strike it too abrupt increase the PF resistance If the heel is too firm or the knee is buckling at heel strike decrease PF resistance Dorsiflexion DF resistance is adjusted to control the advancement of the amputee over the foot...

Страница 4: ...te If the proposed solutions do not result in a satisfactory outcome please contact Freedom Innovations Technical Services for further support 888 818 6777 or 949 672 0032 Spectra Sock A Spectra sock is provided to protect the foot shell and minimize noise Spectra socks must be replaced at intervals appropriate to the user s activity level Failure to inspect and replace the Spectra socks may prema...

Страница 5: ...tist s responsibility to select and or fabricate properly fitting attachment components Never attempt to loosen the bolt affixing the pyramid connector Discontinue use and consult your prosthetist if any part of the prosthesis starts to make noise or the foot loses its responsiveness Inform your prosthetist if you lose or gain a significant amount of weight Avoid use in water as prolonged exposure...

Отзывы: