background image

Instructions for Use (R-720-175, Rev. D)  

 

 

 

 

 

 

 

English

 

Kinterra

 

Instructions for Use 

 

Product Number:  ROM-00-XXXXX-XX

 

 
 
Introduction

 

The Kinterra

 foot/ankle system combines hydraulics and carbon fiber, to provide 

low to moderate impact K3 ambulators an exceptionally normal walking gait 

 regardless 

of surface angle or speed. The result is rock solid stability and the confidence for users to 
choose their own path. 

 

Articulated  ankle  movement  provides  12  degrees  of  motion,  with  adaption  to 
slopes for improved stability and overall comfort. 

 

An  innovative  Dorsi-Assist  Spring  assists  toe  clearance  to  prevent  stumbles  and 
falls. 

 

Precision controls are easily accessible, making the set-up fast and efficient. 

 

The design uniquely delivers gait simulating biological norms on both slopes and 
level ground for superior stability. 

 
Patient Selection

 

The Kinterra

 is intended for use by low to moderate impact transtibial and transfemoral 

K3 level ambulators. It is recommended for unilateral use.  Caution should be used when 
fitting  bilateral patients due to the added movement of the ankle.  Kinterra has a weight 
limit of 275 lbs. (125 kg). 

 
Assembly

 

The Kinterra foot/ankle system is pre-assembled consisting of a foot module, ankle unit, 
Spectra

  sock  and  foot  shell.  After  dynamic  alignment,  torque  pyramid  adjustment 

screws  to 

the  manufacturer’s  specifications.  Secure

  pyramid  adjustment  screws  with  a 

thread locking adhesive (i.e., Loctite 242). 

 

Alignment 

 

 

 

 
 
 
 

Bench Alignment

 

With a 

3/8” (10mm) lift under the heel or the foot placed in the desired shoe, use a plum 

bob or  laser to confirm that the  weight line  falls through the  anterior  aspect of the pylon 
and pyramid.  

 

 

Static Alignment

 

To complete the static alignment, stand the patient in a set of parallel bars.  The patient 
should  be  able  to  stand  comfortably  without  feeling  as  if  the  knee  is  flexing  or 
hyperextending.    If  the  knee  is  flexing,  slide  the  foot  anteriorly.    If  the  knee  is 
hyperextending, slide the foot posteriorly.   

3/8”(10mm)

 

Weight Line

 

 

Содержание Kinterra

Страница 1: ...impact transtibial and transfemoral K3 level ambulators It is recommended for unilateral use Caution should be used when fitting bilateral patients due to the added movement of the ankle Kinterra has a weight limit of 275 lbs 125 kg Assembly The Kinterra foot ankle system is pre assembled consisting of a foot module ankle unit Spectra sock and foot shell After dynamic alignment torque pyramid adju...

Страница 2: ...exion is properly accommodated that the system still has 2 dorsiflexion and 10 plantarflexion Dynamic Alignment Once a stable static alignment is achieved begin walking the patient in the parallel bars to optimize alignment and function on level ground Using your patient s feedback adjust the hydraulic valves carefully to balance the comfort provided by the ankle damper and energy return provided ...

Страница 3: ...ke to mid stance motion before proceeding on to adjust the dorsiflexion adjustment Adjust plantarflexion resistance to dampen heel strike If a foot slap is noticed or heel strike it too abrupt increase the PF resistance If the heel is too firm or the knee is buckling at heel strike decrease PF resistance Dorsiflexion DF resistance is adjusted to control the advancement of the amputee over the foot...

Страница 4: ...te If the proposed solutions do not result in a satisfactory outcome please contact Freedom Innovations Technical Services for further support 888 818 6777 or 949 672 0032 Spectra Sock A Spectra sock is provided to protect the foot shell and minimize noise Spectra socks must be replaced at intervals appropriate to the user s activity level Failure to inspect and replace the Spectra socks may prema...

Страница 5: ...tist s responsibility to select and or fabricate properly fitting attachment components Never attempt to loosen the bolt affixing the pyramid connector Discontinue use and consult your prosthetist if any part of the prosthesis starts to make noise or the foot loses its responsiveness Inform your prosthetist if you lose or gain a significant amount of weight Avoid use in water as prolonged exposure...

Отзывы: