Instructions for Use (R-720-175, Rev. D)  

 

 

 

 

 

 

 

English

 

 

Tip!

  Angular  adjustment  of  the  pyramid  will  impact  the  ratio  of  the  12  degrees  of 

movement.  Plantarflexing  at  the  pyramid  will  decrease  the  dorsiflexion  range  of  motion 
provided by the ankle. Dorsiflexing at the pyramid will increase the plantar  flexion range 
of motion. 
 
Ensure that the range of dorsiflexion and plantarflexion motion is maintained during static 
alignment.  Check  when  flexion  is  properly  accommodated  that  the  system  still  has  2

 

dorsiflexion and 10

 plantarflexion. 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Dynamic Alignment

 

Once a stable static alignment is achieved, begin walking the patient in  the parallel bars 
to optimize alignment and function on level ground. 

.   

Using you

r patient’s feedback, adjust the

 hydraulic valves carefully to balance the comfort 

provided  by  the  ankle  damper  and  energy  return  provided  by  the  foot  module.  (Higher 
resistances  will  allow  the  carbon  fiber  to  be  loaded  more,  provided  more  energy  return.  
Lower resistances will allow more ankle motion and comfort while sitting and ambulating 
on slopes and uneven terrain.)  

Содержание Kinterra

Страница 1: ...impact transtibial and transfemoral K3 level ambulators It is recommended for unilateral use Caution should be used when fitting bilateral patients due to the added movement of the ankle Kinterra has a weight limit of 275 lbs 125 kg Assembly The Kinterra foot ankle system is pre assembled consisting of a foot module ankle unit Spectra sock and foot shell After dynamic alignment torque pyramid adju...

Страница 2: ...exion is properly accommodated that the system still has 2 dorsiflexion and 10 plantarflexion Dynamic Alignment Once a stable static alignment is achieved begin walking the patient in the parallel bars to optimize alignment and function on level ground Using your patient s feedback adjust the hydraulic valves carefully to balance the comfort provided by the ankle damper and energy return provided ...

Страница 3: ...ke to mid stance motion before proceeding on to adjust the dorsiflexion adjustment Adjust plantarflexion resistance to dampen heel strike If a foot slap is noticed or heel strike it too abrupt increase the PF resistance If the heel is too firm or the knee is buckling at heel strike decrease PF resistance Dorsiflexion DF resistance is adjusted to control the advancement of the amputee over the foot...

Страница 4: ...te If the proposed solutions do not result in a satisfactory outcome please contact Freedom Innovations Technical Services for further support 888 818 6777 or 949 672 0032 Spectra Sock A Spectra sock is provided to protect the foot shell and minimize noise Spectra socks must be replaced at intervals appropriate to the user s activity level Failure to inspect and replace the Spectra socks may prema...

Страница 5: ...tist s responsibility to select and or fabricate properly fitting attachment components Never attempt to loosen the bolt affixing the pyramid connector Discontinue use and consult your prosthetist if any part of the prosthesis starts to make noise or the foot loses its responsiveness Inform your prosthetist if you lose or gain a significant amount of weight Avoid use in water as prolonged exposure...