background image

Geiger Counter   523545 

 AC513545

   p.

 4/4 

 

GM voltage 

The tube voltage can be adjusted in increments of 
10 V. Nominal voltage is 500 V. 

It is possible to follow the count rate (

CPS

) in the 

secondary  display  while  the  voltage  is  adjusted. 
Thus,  one  can  fairly  easily  estimate  the  knee 
voltage  of  the  tube.  This  should  be  done  by 

increasing

  the  voltage  as  the  response  time  is 

somewhat longer when the voltage is reduced. 

Press 

CAL

 and adjust the voltage with 

 and 

Select with 

CAL

This  setting  returns  to  default  value  when  the 
instrument is turned off. 

Setting the default value for the GM voltage 

Switch  off  with  OFF.  Keep 

CAL

  pressed  while 

turning  on  again  (until  the  symbol  CAL  is 
displayed).  

Set the voltage with 

 and 

. Select with 

CAL

Statistics 

The instrument keeps track of the maximum, 
minimum and average value of the 

CPS

 and 

µSv/h

. The statistics functions start when the 

instrument is switched on. 
 

The values are displayed like this: 

HOLD 

 

Display: [

HOLD

 

MAX

shows max. 

CPS

 and 

µSv/h 

HOLD

 = exit 

HOLD 



 

Display: [

HOLD

                

MIN

 

Shows min. 

CPS

 and 

µSv/h 

HOLD

 = exit 

HOLD 



 

Display: [

HOLD

        

AVG

 

Shows avg. 

CPS

 and 

µSv/h 

HOLD

 = exit 

 
Resetting 

Instead of pressing 

HOLD

 in the end, press 

CAL

 to 

restart the statistics functions and use the current 
values for 

CPS

 and 

µSv/h

 as new starting point. 

 

The instrument has a USB-like socket placed under 
a cover. This is 

not

 for connection of USB devices. 

(It is used during production for test purposes.) 