TROUBLE SHOOTING
FAULT
POSSIBLE CAUSE
REMEDY
Cabinet does not
operate/start
The mains isolating switch on
the wall, circuit breaker or fuses
are off at the power board
Turn isolating switch circuit
breaker or fuses on
The power switch on the
controller is OFF
Press the power button
The internal fuse has blown
Have circuit checked and
replace fuse
The controller is faulty
Have the controller replaced
HA2 Alarm Display on
Controller
Compressor discharge pipe high
temperature
Turn power off, clean condenser
radiator & pre-filter.
Turn power on.
If no cooling, or alarm returns
within 24 h, call for service.
Refrigerated Cabinet does
not reach temperature
Ventilation grills are blocked
Vacuum or remove blockage
Product blocking air grill
Place product on shelves
Thermostat needs adjustment
Adjust refrigeration controller
Ambient > 25
C & 60%RH
Adjust store air conditioning
Evaporator coil iced up
De-ice coil, check defrost
parameters, replace controller
if found faulty
Condenser radiator blocked
Remove dust and debris
Refrigeration controller faulty
Replace controller
Temperature probe damaged
Replace temperature probe
Fans not operating
Have fans checked/replaced
Condensate Overflows
Defrost cycle unsuitable
Have defrost cycle adjusted
Boil off element failed
Replace element
Cabinet lights not working
LED strip has failed
Replace LED strip
LED power supply has failed
Replace LED power supply
Internal fuse has blown
Have circuit checked and
replace fuse
Aluminium parts corroded
Caustic detergent damage
Order replacement parts
Service
Personnel Only
The table entries in
italics
indicate actions to be taken only by qualified Service
Personnel.