33
FX-300-REL
Alarm Relay and Controller Module
•
FX-300-REL is a versatile controller and alarm module with 2 independent limits
•
Performs alarm relay and/or controller function for FX-300-pH, FX-300-ISE or FX-300-CON
measurement modules
•
Offering Simple On/Off as well as more sophisticated Time Proportional Control (TPC) and
Proportional Frequency Control (PFC) a.k.a. Variable Pulse
•
One FX-300-REL module is required for control and/or alarm function for each FX-300 measuring
module (i.e., 1 each analog input per FX-300-REL module)
•
Analog input: 0-20mA or 4-20mA
•
Maximum or minimum limits can be configured for each of the 2 relays
•
Configurable start timer and reaction timers
•
24VDC power operation, 5A max load/rating for each dry contact relay
•
Hold function to disable relays during calibration and maintenance of measureming transmitters
and sensors
•
Scaling in native measurement units (pH, ppm, μS/mS) for all configurable parameters for ease of
use and programming
•
DIN rail mountable; small form factor for tight space installations
•
Field installations using NEMA 4X enclosures, supporting up to a total of 8ea FX-300 transmitter
modules in a single enclosure.
Application
FX-300-REL modules are ideal for supervising mA signals. This unit is fully compatible with the output
provided by all of our FX-300 measuring transmitters (FX-300-pH, FX-300-ISE, and FX-300-CON).
Simple On/Off as well as programmable control functionality is possible with the FX-300-REL module.
Analog Input
The analog input is a current input, and may be set up to either 0-20mA or 4-20mA. The FX-300-REL
can be scaled so as to match exactly the 0-20mA or 4-20mA analog output from the FX-300-pH, FX-
300-ISE & FX-300-CON measurement modules.
Digital Input
The FX-300-REL may be configured to both positive and negative logic on the digital inputs. Using
parameter no. 21, the input may be set to either active high or active low. With positive logic, the
accepted input voltage range is 5- 30VDC. Negative logic is achieved by connecting the input to
ground; for example by using a switch. The supervision may be blocked by activating the S2 input.
Relay Output
The unit contains two relays, one for each limit. The relays are both connection relays, but the polarity
may be inverted independently using parameters no. 19 and 20 when in simple supervision and On/
Off control mode.