32
Fluoride Sensor Specifications
FX-300 Integrated Fluoride Ion Selective Sensor Specifications Twist Lock Mounting
Part
Number:
SF-8T-UL-10
1" MNPT Twist Lock Immersion Integrated, Fluoride Ion Selective Sensor for use in high pH Applications
Recommended Applications:
Fluoride ion concentration in aqueous solution from drinking water
through waste water. Not suitable for use in etching
solutions.
Special Features:
Crosslinked polymer in the reference system is resistant to heat, solvents and to
most chemicals. Sensor holds an excess of KCl, assuring saturation at all temperatures and extending
the life of the
sensor.
The construction of the sensor permits easy access to the sensing and reference surfaces for
cleaning or inspection.
General Specifications:
Concentration
Range:
1 to 10
-6
Molar, 19,000 to
0.019 ppm
Lowest Limit of Detection
5 X1 0
-8
Molar, .001 ppm
pH Range:
5.5 to 9.5 pH (continuous), intermittent up to 11 pH with measuring crystal degradation
Temperature
Range:
5 to 50
o
C
Temperature Compensation:
Fast Temp PT1000 for
variable temperature
conditions
Pressure
Range:
1 to 20 psig (6.9 to 138 kPag)
Body
Material:
Ultem (Poly-Ether-Imide)
Junction
Material:
Kynar (Poly-Vinylidene-Fluoride)
Cable: RG 174/U Coaxial (without preamplifier)
Connector: BNC (unless otherwise
specified)
Ion Sensor Specifications:
Measuring Membrane:
Selective Fluoride Sensitive
Membrane (solid state)
Dimensions:
0.310", (7.8 mm) DIA
Initial
Impedance:
Less than 100 M Ohms @ 25
o
C
Interfering Ions in Ratios of Permissible Excess:
Interfering Ion / Measured Ion (in
Molarity):
OH- above 12.0
pH
Reference System Specifications:
Type: Double
Junction
Reference Half Cell: Ag/AgCl, Saturated
KCl
Primary Junction: Porous Ceramic, Saturated KCl in
crosslinked polymer
Secondary Junction: Porous Kynar, Saturated with KCl
in crosslinked polymer
Surface Area:
366,000 mil
2
, (236 mm
2
)
Storage and Shelf Life: 1 year from date of manufacture at room temperature with protective cap on
7 Sensor Specifications
Содержание FX-300-F
Страница 2: ...Page Intentionally Left Blank...
Страница 6: ...3 2 1 HF Ionization 2 Introduction...
Страница 22: ...19 3 63 Installation FX 300 FP Inline Flow Panel...