FX-300-ISE Module Description
2.45 Default Ion Settings
Your FX-300-ISE has been preconfigured at the factory for your ion selective measurement
requested at the time of order. The ISE measurement type configured for the FX-300-ISE transmitter
cannot be modified in the field.
Below are the default nominal values for the parameter P14 and P15 for the fluoride ion selective
measurement.
Slope Fluoride Ion (Parameter P15 on FX-300-F): -57.2 mV per decade default value
Offset Fluoride Ion (Parameter P14 on FX-300-F): -47 mV default value
Formula Weight of Fluoride Ion (parameter P13 on FX-300-F): 19.00 grams per mol. This is a display
only value that clearly denotes the ISE measurement type.
The slope parameter will only be changed when a 2-point slope calibration is performed. After
calibration the slope should be at or very close to the default value. If not this indicates the calibration
was not performed properly. Parameter P 15 allows you to both view and manually modify the
working slope.
2.46 Transmitter Specifications
Housing
Levan UL94V-0 (Upper part)
Noryl UL94V-0 (Lower part)
Mounting
M36 for 35mm DIN rail
Dimensions
D 58 x W 36 x H 86 mm (2.3 x 1.4 x 3.4 in)
Temperature range
-15 to +50°C
Power supply
24VDC +/- 10%
Consumption
60mA max
Sensor type
Combination sensor
ISE/mV range
0-10, 0-100, 0-999 ppm; +/- 1000mV
ISE input
<1pA, >10GΩ
Accuracy
+/- 2%
Temp sensor
Pt100, Pt1000
Temp range
0-150°C +/- 0.3°C
Temperature Compensation
Fixed (manual) or Automatic (using Temperature (TC)
measurement)
Analog output
0-20mA or 4-20mA, max 250Ω
Enclosure
Wall mount windowed NEMA 4X reinforced fiberglass
2 Introduction