After unpacking, it is recommended to save the shipping carton and packing materials if
the instrument must be stored or re-shipped. Inspect the equipment and packing materials
for signs of shipping damage. If there is any evidence of damage, notify the transit carrier
immediately.
The shipping container consists of the following:
1
FX-1000p Amperometric Chlorine Analyzer
1
Buffer Bottle Bracket
1
Sample Tubing
1/4” id x 3/8” od x 3’ length, clear flexible PVC tubing
1
Sample Drain Hose 5/8” id x 3/4” od x 3’ length, clear flexible PVC tubing
1
Overflow Drain Hose 5/8” id x 3/4” od x 3’ length, clear flexible PVC tubing
2
Buffer Pump Tubing #13 neoprene tubing x 9”
1
Set of Four Mounting Feet with Four Screws
1
Instruction Manual
Serial Number
The FX-1000p Amperometric Chlorine Analyzer measurement system has two serial number labels
found on the bottom right side of the enclosure, and on the top of the base plate inside the enclosure.
The label on the outside of the enclosure indicates the chlorine residual range. Should technical
assistance be required, refer to the serial number to identify your system.
Mechanical Requirements
Location
Mount the FX-1000p Amperometric Chlorine Analyzer in a location that is:
• Out of direct sunlight
• Protected from the elements (rain, snow, ice, dust, etc.)
• As close to the sample point as possible to minimize sample lag time
• Accessible
• Clean and dry where there is little or no vibration
• Protected from corrosive fluids
• Within ambient temperature limits (32-120°F, 0-49°C)
Important: Mounting the system in direct sunlight may increase the ambient temperature above the
maximum specified limit.
Mounting
1. Attach the four mounting feet to the rear of the electronics enclosure using the included flat head
screws.
2. Orient the analyzer so that the sample flow cell drain outlets are pointing downward.
3. Anchor securely to a wall at approximate shoulder height for convenience.
Installation
8