Lysis Buffer
Elution Buffer
Wash Buffer
Wash Buffer
Wash Buffer
Magnetic Beads
Lysis Buffer
Elution Buffer
Wash Buffer
Wash Buffer
Wash Buffer
Magnetic Beads
Load Liquid Sample
Load Liquid
Sample
Load Liquid
Sample
Collect and transfer
magnetic beads to
lysis buffer column
Sample lysis and binding of
nucleic acids to magnetic beads
Transfer magnetic beads to
wash buffer column
Magnetic Rods
Magnetic Rod Sleeves
Deep-well Plate
Stir up
and down
1 2
3 4 5 6 7 8 9 10 11 12
Stir up and down
1 2
3 4 5 6 7 8 9 10 11 12
1 2
3 4 5 6 7 8 9 10 11 12
Lysis/Nucleic Acid Binding:
Lysis buffer solution promotes cell rupture and releases
nucleic acids. Magnetic beads transferred from wells 2/8 to the lysis wells binds to the
target nucleic acid molecules. The magnetic beads with the bound particles are then
moved to the wells 3/9 for washing.
32 samples per run. The automated system integrates lysis, binding,
mixing, washing, and elution steps to extract nucleic acid of high purity.
Magnetic beads, with strong affinity and adsorption force for nucleic acids,
binds to the nucleic acid released on lysis. The nucleic acids bound to the
magnetic beads can be transferred to different wells using the magnetic
rods. The magnetic rod sleeves facilitate quick and even mixing of the
buffer and the nucleic acid bound beads.
Working Principle and Steps
Sample Loading
Remove the Sealing Film
3