6.2. Dual point calibration
Due to the influence of ambient temperature, humidity, and pot
structure, the water bath has a certain deviation in temperature
measurement. In order to improve the measurement and control
accuracy, each single instrument is calibrated before shipment.
Includes two calibration options -- low temperature point and high
temperature point.
Low temperature point(LCP): 40°C
High temperature point(HCP): 95°C
Low temperature point calibration:
For example:
Before calibration, the screen displays 40.0 C but the actual liquid
temperature is 42
.
0°C
After calibration, the screen displays 41.8 C and the actual liquid
temperature is 42
.
0°C
High temperature point calibration:
For example:
Before calibration, the screen displays 95°C but the actual liquid
temperature is 98.0°C;
After calibration, the screen displays 97.7°C and the actual liquid
temperature is 98.0°C;
Operation
Display
Select the menu option
Select the low temperature point calibration
setting menu option
Press SET to enter the calibration value
Enter the reference thermometer value
Press to store the calibration value, or the
setting is invalid. The setting will exit
automatically after 10S without operation
13
Содержание Ironman III
Страница 1: ...Operating Instructions Water Bath WB301 Ironman...
Страница 17: ...1 2 3 3 1 3 2 3 3 4 4 1 4 2 4 3 5 6 7 8 16 17 17 18 18 20 20 20 20 20 20 21 25 27 27 27 28...
Страница 18: ...17 2014 35 EU EN61010 1 2014 30 EU EN55011 B EN61326 1 YY91037 99...
Страница 19: ...18 1 2 Ironman 1 1 1 1 1...
Страница 20: ...19 3 3 1 3 2...
Страница 21: ...3 3 4 4 1 4 1 1 4 1 2 1 20...
Страница 22: ...2 4 1 3 60 4 1 4 4 2 4 2 1 21 C F...
Страница 23: ...22 4 2 3 50 0 37 0 55 0 37 0 1 4 2 4 50 0 37 0 OFF OFF 50 0 OFF OFF 4 2 2 OFF 50 0...
Страница 24: ...23 OFF 1 4 2 4 3 P1 50 0 37 0 P1 4 2 5 OFF OFF...
Страница 25: ...24 4 2 7 0 99 59 3 OFF 1 2 3 4 0 1 10 OFF...
Страница 27: ...26 5s 95 98 0 97 7 98 0 5s...
Страница 29: ...8 2 28 3 7 915013000...