Four E's Scientific CF0201002 Скачать руководство пользователя страница 6

 

Error 4     

Over speed   

Power off and re-start at lower speed 

Error5     

Abnormal vibration         

Load the rotor again 

Load the tubes Symmetrically 

Tighten the rotor lid firmly 

Error 6 

Door failure 

Power off and power on 

Use the screw driver into the door 

release hole

 

 

 

Open the door

 

 

(1) If powered on 

 

 

The door will open automatically after the running completed. 

 

Open the door by pressing the button 

“Door” 

 

(2) If power outage 

   

 

The  door  cannot  open  automatically  if  there

’s  a  power  outage.  It 

can be opened manually. 

 

Make sure the running stops completely 

 

Insert a screw driver into the hole on the top left side of the unit and 
push forward to open the door.   

 

 

Accessory

 

 

 

1* rotor (1.5mL/2mLx 12)

 

 

1* rotor locking nut

 

 

1 * rotor lid

 

 

1* wrench

 

 

1* screw driver (to open the door)

 

 

1* power adapter

       

 

 

 

Specifications 

 

Items 

CF0201002 

Maximum Speed (rpm)

 

15000rpm  (500-15000rpm), 

increment:100rpm

 

Maximum RCF (g)

 

15100 x g, increment: 100 x g

 

Speed Accuracy

 

±20rpm

 

Rotor Capacity

 

0.2ml/0.5ml/1.5ml/2ml×12

 

Time Range

 

30sec-99min-HOLD 

(Continuous Operation)

 

Driving motor

 

Brushless DC motor

 

Safety devices

 

Door  interlock;  Over-speed 

detector; 

Over-temperature 

detector;  Automatic  Internal 

Diagnosis 

Power

 

Single-phase,   

AC 100V-240V, 

Содержание CF0201002

Страница 1: ...ctions Safety Instructions 2 Installation 2 Structure 3 Operation Panel 3 Rotor Loading 4 Operation 4 Maintenance 5 Trouble Shooting 5 Open the door 6 Accessory 6 Specifications 6 Warranty 7 High Spee...

Страница 2: ...tor is rotating never release the door lock Unauthorized repairs disassembly or modifying the centrifuge are strictly prohibited The machine must be placed on a firm and flat table Make sure the centr...

Страница 3: ...ick spins The centrifuge will keep running by pressing the button and stop running by releasing the button It only works when the unit stops and cannot work when the unit is running Select and save th...

Страница 4: ...et is 500 15000rpm and the increment is 100rpm Rotate the speed knob to set the speed value Press the speed knob to confirm the speed value Press the speed knob twice RPM will change to RCF and the RC...

Страница 5: ...ure Clean the centrifuge with cloth or sponge moistened with a neutral detergent solution Sterilize the centrifuge with cloth moistened with 70 ethanol solution Do not directly pour water neutral dete...

Страница 6: ...d manually Make sure the running stops completely Insert a screw driver into the hole on the top left side of the unit and push forward to open the door Accessory 1 rotor 1 5mL 2mLx 12 1 rotor locking...

Страница 7: ...elow cases Failures caused by incorrect installation Failures caused by rough or improper handling Failures caused by conveyance or relocation after installation Failures caused by unauthorized disass...

Отзывы: