Rotor Loading 

 

 

 

Load the supplied rotor onto the shaft and make sure the rotor is in 

position and connected well with the shaft. 

 

Hold the rotor with one hand and tighten the nut with the supplied 
wrench firmly (refer to the below picture) 

 

Check the rotor is firmly tightened before running. 

 

Put on the rotor lid 

 

 

Close the door and make sure the door is closed completely. 

 

Operation 

 

 

(1) Set the speed RPM/RCF 

 

 

Press the speed knob once and the speed value will flash 

 

The speed range can be set is 500-15000rpm and the increment is 
100rpm. 

 

Rotate the speed knob to set the speed value 

 

Press the speed knob to confirm the speed value 

 

Press the speed knob twice, RPM will change to RCF and the RCF 
value will flash. Rotate and press it to set and confirm the RCF. 

 

After set the speed without pressing the button for confirmation, the 
system  will  stop  the  value  flash  and  confirm  the  set  value  after 

several seconds. 

 

(2) Set the time 

 

 

Press the time knob and the time value will flash 

 

The time can be set from 30 seconds 

– 99 minutes 

 

Rotate the time knob to set the time 

 

 

Press the time knob to confirm the time value 

 

Symmetrically load centrifuge tubes into the rotor to keep the balance.     

 

Содержание CF0201002

Страница 1: ...ctions Safety Instructions 2 Installation 2 Structure 3 Operation Panel 3 Rotor Loading 4 Operation 4 Maintenance 5 Trouble Shooting 5 Open the door 6 Accessory 6 Specifications 6 Warranty 7 High Spee...

Страница 2: ...tor is rotating never release the door lock Unauthorized repairs disassembly or modifying the centrifuge are strictly prohibited The machine must be placed on a firm and flat table Make sure the centr...

Страница 3: ...ick spins The centrifuge will keep running by pressing the button and stop running by releasing the button It only works when the unit stops and cannot work when the unit is running Select and save th...

Страница 4: ...et is 500 15000rpm and the increment is 100rpm Rotate the speed knob to set the speed value Press the speed knob to confirm the speed value Press the speed knob twice RPM will change to RCF and the RC...

Страница 5: ...ure Clean the centrifuge with cloth or sponge moistened with a neutral detergent solution Sterilize the centrifuge with cloth moistened with 70 ethanol solution Do not directly pour water neutral dete...

Страница 6: ...d manually Make sure the running stops completely Insert a screw driver into the hole on the top left side of the unit and push forward to open the door Accessory 1 rotor 1 5mL 2mLx 12 1 rotor locking...

Страница 7: ...elow cases Failures caused by incorrect installation Failures caused by rough or improper handling Failures caused by conveyance or relocation after installation Failures caused by unauthorized disass...

Отзывы: