Model 700 Series _____________________________________________Maintenance
4-1
Section 4 - Routine Maintenance
4.1 Cleaning The Cabinet Exterior
1.
Wipe down the freezer exterior using soap and water and a general use laboratory
disinfectant.
2.
Rinse thoroughly with clean water and dry with a soft cloth.
4.2 Cleaning the Air Filter (minimum of twice a year*)
1.
Locate the grille on the right side of control panel. See Figure 1-2 or 1-3. Grasp the
corner of the grille and gently pull to remove.
2.
Remove the filter material and wash, using water and mild detergent.
3.
Dry by pressing between two towels.
4.
Install the filter back into the grille and attach the grille.
* Depending upon environmental conditions, the filter may need to be cleaned or replaced more
frequently. If the filter becomes torn or excessively dirty, a replacement can be purchased from
Forma Scientific. See the Parts List, Section 7, for filter part number. A filter kit (set of 5 filters)
is also available.
4.3 Cleaning the Condenser (minimum of twice a year*)
1.
Locate the grille on the right side of control panel. See Figure 1-2 or 1-3. Grasp the
corner of the grille and pull gently to remove.
2.
Using a vacuum cleaner, exercising care to not damage the condenser fins, clean the
condenser.
3.
Install the grille.
* Depending upon environmental conditions, the condenser may need to be cleaned more
frequently.
Avoid the excessive use of water around the control area due
to the risk of electrical shock. Damage to the controls may
also result.