Forma
Model 700 Series____________________________________Installation/Start-Up
1-7
1.9 Attaching the Power Cord
1.
Loosen screw (A) located on the power
cord retainer.
2.
Spread the power cord retainer.
3.
Insert the power cord into the power outlet
module (B).
4.
Tighten screw (A) on the power cord
retainer.
1.10 Power Switch (mains disconnect)
The power switch is located on the rear of the unit, directly above the power cord.
The power switch is also a circuit breaker that protects the entire unit.
1.11 Electrical Requirement and Connection
•
The freezer should be operated on a dedicated grounded service.
•
Check voltage rating on the serial tag of the unit and compare it with the outlet voltage.
•
With the power switch turned off, plug the line cord into the wall outlet.
1.12 Factory Settings
•
Control Temperature – 707 & 771: -80°C
– 733 & 754: -75°C
•
Over Temperature Alarm: -65°C
If you wish to change any of these settings, see the appropriate section(s) in Section 2.
1.13 Start Up and Loading
•
Turn the freezer on and allow it to run
empty
overnight.
•
When the empty freezer has stabilized overnight at the control temperature set point, load
the chamber with pre-frozen product.
Figure 1-9
Power Cord Assembly
Make sure the power cord connection is completely seated.
The freezer was designed for the storage of pre-frozen product
only. The addition of warm product may cause a temporary rise in
the cabinet temperature.