47
Error Messages
MESSAGE
WHAT IT
MEANS
WHAT DOES
METER SAY?
WHAT TO DO
Appears when
the
batteries cannot
provide enough
power for a test.
Battery is dead.
Please replace.
Replace the bat-
teries immedi-
ately.
Remove the
strip after apply-
ing blood to the
absorbent hole.
Test strip has
been removed
during testing.
Appears when a
used test strip
is inserted.
The test strip
has been used.
Please replace.
Re-test with a
new test strip.
Repeat with a
new test strip.
Appears when
ambient
temperature is
below system
operation range.
System op-
eration range is
50°F to 104°F
(10°C to 40°C).
Repeat the test
after the
meter and test
strip are in the
above tempera-
ture range.
Room tem-
perature is
out of range.
Thus unable to
measure.
Appears when
ambient
temperature is
above system
operation range.