23
Service Manual
båÖäáëÜ
cê~å´~áë
3 Troubleshooting
Problem
Possible solution
Unit cannot be turned
on.
● Check whether the power cable is plugged
in.
● Press the unit's ON/OFF switch.
● If the problem persists, disconnect the unit
from the power supply and check the fuses.
The camera heater
does not heat.
● Check whether the camera is connected
correctly.
● Check the heater's fuse.
The fuse has blown.
NOTICE! Replacing the fuse may cause
damage to the unit.
● Switch the unit off and disconnect it from
the power supply.
● Have the fuse replaced by trained,
specialist personnel or a service engineer.
The exposure cannot
be performed.
● Check whether the software is displaying
the ACQUISITION phase.
● Check whether the camera is switched on
and the camera is displaying a live image.
● Check whether the camera is indicating an
error.
● Check whether the teeth are covered with
MyCrown HD in line with the instructions.
No connection can be
established to the
internet.
● Check the connection to the wireless
network.
● Check the Internet connection.
● Restore the connection to the wireless
network if necessary (see operating
instructions).
● Briefly disconnect the wireless network
router from the power supply. Restart
MyCrown Design.
Содержание MyCrown Design
Страница 1: ...j a p j b Service manual...
Страница 21: ...Service Manual b c 21 2 6 Attaching the cover 1 Attach the back cover on the column 2 Connect the camera cable...
Страница 70: ...70 MyCrown Design 4 Attach the rear panel 5 Attach the back cover...
Страница 72: ...72 MyCrown Design 3 Remove the LAN cable from the respective coupling on both sides...
Страница 74: ...74 MyCrown Design 3 Attach the lower cover 4 Mount the upper cover...
Страница 76: ...76 MyCrown Design 4 Remove the strain relief 5 Carefully pull the cable plug upwards...