![Foerster 6.461.21 Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/foerster/6-461-21/6-461-21_operating-instructions-manual_2314653045.webp)
SENSORSY STEM Ro 35 L
OPERATION
4.4 Rotational speed preselection
6.461.21 - 2008/06
4-13
4.3.5 Adjusting roller guide
Open roller guide in accordance with material diameter.
Insert test material.
Adjust roller guide so that the rollers just touch the test material.
NOTE
!
The test piece must run centrically through the sensor system, if nec-
essary adjust the table to the height of the test axis and lock.
4.4 Rotational speed preselection
The following rotational speeds are available depending on the type of
motor control:
MOC E motor control
fixed rotational speed of 4,500 rpm
50 Hz mains supply
MOC S/SB motor control
Rotational speed switchable 4,500 or 9,000 rpm,
with or without electrical brake system
Both motor controls supply the required voltages and currents and contain
the necessary switching and safety devices (contactors, motor protection
switches, protective circuits).