SENSOR SYSTEM Ro 35 L
4.3 Adjusting nominal diameter
OPERATION
4-8
6.461.21 - 2008/06
4.3.3 Changing
nozzles
Unscrew the nozzles from the nozzle holder using two hook spanners.
CAUTION
!
If the nozzle holder is fixed in a vice, clamp on the surfaces
only!
Screw in new protective nozzles.
Fix nozzles using hook spanners.
Check end position! Danger of collision with test head!
NOTE
!
Use only nozzle pairs with the same nominal diameter!
WARNING
!
Adjust the test heads to the nominal diameter of the protective
nozzles before inserting the nozzle holders!
Inserting the nozzle holders into a test head arrangement
that is too narrow can cause damage to the test
heads and/or nozzles.
Fig. 4.7 Nozzle holders, nozzles and hook spanners
Outside nozzle holder
Hook spanner
Outside protective nozzle
Inside protective nozzle
Inside nozzle holder