7
the measuring beaker, and place the beaker inside the machine in its fitting.
Make sure the pipes inside the machine are in
proper position!!
T
he premounted silicon hoses on
the pipes should be in contact with the pipeholder,
and in that way guarantee correct positions,
but that
may need to be fine-tuned by the user,
so take
very
good care
to ensure the following pipe-positions (fig
2a,2b):
Wait to install the pH-probe until it is calibrated, let
it be in the transport cup until then.
•
Pipe from Pump C should be a few cm above surface of the water.
•
Pipe from pump B must be in contact with the bottom. Very important!!
•
Pipe from pump A must be around 2 mm above bottom. Very important!!
Establish software connection
The next step is to establish a connection between the APP in the phone/tablet and
the machine.
General information:
The connection between Alkatronic and App concerning the most crucial
functions, is based on BT-connection, as that is the most safe and stable
connection. But it could take sometimes 5-10 seconds for the app to connect
when opening the app from a sleeping mode. That is normal!. Just wait, and
Figure 2a
Pump A
Pump C
Pump B
Pump D
Figure 2b