iTuner Manual
32
Examples:
TERM 2 0.15 125
Defines termination for first secondary frequency
4.4.10
TuneTo Command
Call TUNETO to set reflection coefficient in DUT reference plane. The tuner will move the probe
to the new position.
Syntax:
TUNETO mag [phase]
Parameter:
Mag
Phase
Magnitude of reflection coefficient
Phase of reflection coefficient (optional)
Return value:
0 -> okay
-1 -> no calibration loaded
Remarks:
Phase parameter is optional. If omitted, tuner only adjusts vertical axis to achieve
desired magnitude.
Use TuneTo? to determine the new probe position without moving the probe
Use STATUS? To detect completion of axis movement.
Use GAMMA? To read back actual reflection coefficient in DUT reference plane
Examples:
Tuneto .5 90
Tuneto .8
Tunes to mag=0.5 phase=90 degrees
Sets mag=0.8 for current horizontal position of probe
4.4.11
TuneTo? Command
TUNETO? Allows to query the probe position of the tuner for the specified gamma without
actually moving the probe.
Syntax:
TUNETO? mag [phase]
Parameter:
Mag
Phase
Magnitude of reflection coefficient
Phase of reflection coefficient (optional)
Return value:
horizontal and vertical position of the probe for the specified goal