Bolt elbow fitting

Set screws elbow fitting

The elbow fitting can be adjusted in two directions:

Adjusting the rotation of the elbow fitting: loosen the “bolt elbow fitting” until the fitting 

can be rotated. Adjust the fitting in the desired position and fasten the “bolt elbow fitting”.

Adjusting the position of the elbow fitting: loosen one or multiple set screws of the elbow 

fitting. Adjust the elbow fitting by shifting this fitting in the direction of the arrow. Fasten all 

set screws of the elbow fitting.

Adjust the balancing force 

 

Warning: 

Adjusting the balancing force can result in forces that are higher than the mass 

of the arm, resulting in an upwards movement of the arm.

The upwards balancing force can be adjusted by turning the adjustment knob. The force 

required to rotate the adjustment knob can change depending on the position of the human 

interface. This is normal behaviour. Because of this, the system can be adjusted using little 

force. Rotating the knob clockwise will give more balancing force, counter clock wise will give 

less force.

 

At the top of Dowing

2

 there are two eject knobs. Pressing these two towards each other 

makes it possible to remove the human interface, containing the loadarm, arm fitting, elbow 

fitting and wrist support.

 

Body of Dowing

2

. The body of Dowing

2

 contains: the balancing system including the 

possibility to adjust the balancing force. 

 

Balancing force adjustment knob. 

 

Swivel arms 

 

Level rotation head

 

Locking hook

 

Table clamp

Adjusting the arm fitting and elbow fitting

To move the arm freely the weight of the arm is balanced. To balance the arm in the vertical 

direction, considerable forces are required. For safe use it is important that: 

 

the arm is positioned in a stable way in the arm fitting

 

the arm is always in contact with the elbow rest

If you notice that the arm is not stable and tends to slip out, reposition the arm in the proper way. 

 

 

Warning: 

The positions of arm fitting and elbow fitting are crucial for the performance 

of Dowing

2

. Changing these positions can result in a severe decrease of the performance 

or even malfunction of Dowing

2

. Therefore only trained persons are allowed to change 

the settings of the arm fitting and elbow fitting.

 

Warning: 

Adjusting the arm fitting can result in malfunction of Dowing

2

Set screws arm fitting

The arm fitting can be adjusted in one axis. Adjusting the position of the arm fitting is done in 

the following way: Loosen one or multiple set screws of the arm fitting. Adjust the arm fitting 

by shifting this fitting in the direction of the red arrow. Fasten all set screws of the arm fitting. 

Содержание Dowing2

Страница 1: ...ID manual 606281 Revision 1 0 0 Date 1 May 2020 User manual Dowing2...

Страница 2: ...trademark of Focal Meditech BV Dowing2 is a registered trademark of Focal Meditech BV Dowing2 Model Design is registered by Focal Meditech BV Content Disclaimer 2 Copyright Notice 2 Trademarks 2 Cont...

Страница 3: ...essential information for using Dowing2 information about safety issues and contact information Please read this information carefully increase of knowledge of the arm support will result in an increa...

Страница 4: ...aution This symbol in combination with the word Caution is used to warn about possible unsafe practices Extra attention is required Disposal This symbol indicates that this product is not to be dispos...

Страница 5: ...ical energy is stored for balancing the arm During non intended removal of the arm from the arm fitting of Dowing2 this mechanical energy will be released resulting in a fast moving human interface wh...

Страница 6: ...aluation attention is required to determine if the user is able to sit in a stable position and if one can stand up without using a support Intended use and operation of the device Intended use Dowing...

Страница 7: ...all intended to be used by persons challenged by limited muscle force in their arms and shoulder girdle Due to diminished use of their musculoskeletal functions prior to the supply of Dowing2 and als...

Страница 8: ...ontaining the loadarm arm fitting elbow fitting and wrist support Body of Dowing2 The body of Dowing2 contains the balancing system including the possibility to adjust the balancing force Balancing fo...

Страница 9: ...itting STEP 2 Apply a small force downwards and slide the locking ring forwards over the hook Disconnecting the human interface Danger Please carefully disconnect remove the different parts and take n...

Страница 10: ...g2 STEP 2 Remove the user s arm STEP 3 Remove Dowing2 from the parking position Warning the arm fitting will move upwards due to the energy stored in the spring STEP 4 Remove the human interface See d...

Страница 11: ...r the user to make a straight line To change the angle loosen the bolt first before adjust the locking hook If the locking hook is set to the desired angle tighten the bolt with 6 Nm Edge depth Edge d...

Страница 12: ...over the tube and tighten the screws For a proper installation the tube should slide at least 25 mm into the bracket Tighten both screws with 6 Nm Adjusting the angle of the rotation head Dowing2 will...

Страница 13: ...lifetime of system parts and interfere with operation All housings must be regularly inspected If any housing is visibly damaged do not use the device It is prohibited to physical modify Dowing2 Ther...

Страница 14: ...0 45 45 Dowing2 dimensions human interface excluded Dimensions of the Dowing2 are variable This is because it is a device with moving parts and different configurations All dimensions are given in mm...

Страница 15: ...03145 Elbow fitting right X Large 603140 Elbow fitting left XX Large 603146 Elbow fitting right XX Large 603109 Arm fitting size 1 left 603111 Arm fitting size 1 right 603113 Arm fitting size 2 left 6...

Страница 16: ...ght Small 603135 Elbow fitting left Large 603141 Elbow fitting right Large 603139 Elbow fitting left X Large 603145 Elbow fitting right X Large 603140 Elbow fitting left XX Large 603146 Elbow fitting...

Страница 17: ...ize 5 right 605658 Inlay arm fitting size 1 left Neoprene 605638 Inlay arm fitting size 1 right 605662 Inlay arm fitting size 2 left 605642 Inlay arm fitting size 2 right 605666 Inlay arm fitting size...

Страница 18: ...ick start guide contains the basic information regarding Dowing2 The aim of this information is to ensure successful safe and effective use of the device This quick start guide will give you a quick l...

Страница 19: ...leiden tot onjuist functioneren van het apparaat of tot lichamelijke schade Danger Prevent extreme temperature see environment conditions in user manual Failure of this can lead to malfunctioning of...

Страница 20: ...t using Dowing2 Parkeer of start gebruik Dowing2 If Dowing2 is not used for a short time it is recommended to place the Dowing2 in its intended parking bracket This will provide a safe situation for b...

Страница 21: ...tkoppelen van de Human Interface Figure 4 Attach detach human interface Figuur 4 Koppelen ontkoppelen human interface The human interface including the loadarm arm fitting elbow fitting and wrist supp...

Страница 22: ...one hand the body of Dowing2 with the other hand lift up in a straight line the lower swivel arm from the clamp STEP 6 Store Dowing2 in its case Dowing2 loskoppelen STAP 1 Vergrendel Dowing2 in de ve...

Страница 23: ...a 31 0 13 533 31 03 en we zenden u een geprint exemplaar toe 12 Adjust the balancing force Aanpassen balanceerkracht Figure 7 Adjust the balancing force Figuur 7 Aanpassen balanceerkracht The upwards...

Страница 24: ...46 47 Appendix 5 Declaration of conformity 14...

Страница 25: ...Focal Meditech are subject to conditions of the national or local representative and in accordance with national law Conditions and Warranty direct supply by Focal Meditech BV to consumers In the case...

Страница 26: ...50...

Отзывы: