Operation Manual EMGZ 490 

 

 

15.12.2015

 

 

23 

 
 

The individual

 

FBs

 

read

 

or

 

write their data

 

to data 

blocks

, which are 

listed 

below

 

 
 
 
 
 
 
 
 
 

 Fig. 

6: 

 Functional Blocks FB1 – FB2  

E490012

 

 
 

7.5

 

Data Blocks 

In a controller, data blocks can be configured. As FBs they can be easier apply and 
monitored by the computer. Basically data blocks are structures that may contain 
different variables. Below three data blocks (DB1, DB2, DB3) were defined.

 

 

 

DB1 – FMS_ACYCLIC_DB: 

Name 

Type 

Start Value 

Comment 

Index 

INT 

FMS specific index 

VALUE_READ 

DINT 

L#0 

Input value 

VALUE_WRITE DINT L#0 

Output 

value 

 
DB2 – FMS_TRIGGER_DB: 

Name 

Type 

Start Value 

Comment 

READ_TRIGGER 

BOOL 

FALSE 

Trigger acyclic read instruction 

WRITE_TRIGGER 

BOOL 

FALSE 

Trigger acyclic write instruction 

 
DB3 – FMS_CYCLIC_DB: 

Name 

Type 

Start Value 

Comment 

ACT_VAL_ADC 

WORD 

W#16#0 

Feedback in ADC 

ACT_VAL_NEWTON  DWORD 

DW#16#0 

Feedback in Newton 

ACT_VAL_POUND 

DWORD 

DW#16#0 

Feedback in Pound 

ACT_VAL_UNIT 

DWORD 

DW#16#0 

Feedback in configured unit 

STATUS BYTE 

B#16#0 

Operation 

status 

 

 

 

Содержание EMGZ 490

Страница 1: ...ersion V1 16 GSDML file V2 3 This operation manual is also available in German Please contact your local FMS representative Diese Bedienungsanleitung ist auch in Deutsch erh ltlich Bitte kontaktieren...

Страница 2: ...n 9 4 3 Calibration Procedure 10 4 4 Gain 10 4 5 Limit Violations 11 4 6 Description of Signal LEDs 12 5 Integration in a PROFINET network 13 5 1 PROFINET Interface 13 5 2 MAC Address 13 5 3 Assignmen...

Страница 3: ...8 8 3 Offset Compensation and Calibration over the Web Browser 29 8 4 Ethernet Settings 30 8 5 System Settings 31 9 Mechanical Dimensions 32 10 Technical Specification 34 10 1 PROFINET Characteristics...

Страница 4: ...igh risk for persons to get health injury or loss life It has to be followed strictly b Risk of damage of machines Caution This symbol refers to information that if ignored could cause heavy mechanica...

Страница 5: ...erature of the amplifiers that are mounted in a cabinet to a minimum a distance of at least 15mm 0 6 between the devices must be kept During operation the use of not PROFINET related services such as...

Страница 6: ...Description EMGZ 490 The microprocessor controlled amplifiers of EMGZ 490 series are used to amplify process and transmit force sensor signals to devices connected in a suitable network The measured f...

Страница 7: ...on ref to chapters 4 1 to 4 3 4 If required do additional settings chapter 8 2 5 In tegrate the amplifier in the PROFINET network chapter 5 3 3 Installation and Wiring Caution In order to improve the...

Страница 8: ...25mm2 AWG 23 shielded twisted pair cable The cable must be installed separate from power lines Fig 2 Wiring Diagram EMGZ 490 Color scheme IEC60757 and pin codes are valid for FMS components only Cauti...

Страница 9: ...the Gain adjusts the controller and load cells so that the display gives you the actual tension value There are two methods of calibrating the system The first method uses a defined weight The second...

Страница 10: ...prevailing force is not passed one to one to the sensors with the result that the measured material tension does not correspond to the effective force In order to account for this error the measured...

Страница 11: ...ost accurate results 4 5 Limit Violations The amplifier checks the analogue input and output on limit violations It checks whether the force sensor is mechanically overloaded overload test At the outp...

Страница 12: ...w The overflow is thus triggered at the output reaching the theoretically calculated value 10 05V For the underflow applies value of 0 05V Upon reaching one of these limits the corresponding Status Bi...

Страница 13: ...the network card of the corresponding device It is set at the factory for each EMGZ 490 device and subsequently it cannot be changed The MAC address is lost after a firmware update It must be re assig...

Страница 14: ...ce as a FMS product worldwide 5 4 TCP IP Configuration In order to enable data communication between amplifier and a PLC or a web browser the Ethernet settings must be known In a PROFINET network the...

Страница 15: ...mode For normal operation the assignment of the address over the PLC is required 5 5 System Start up At system start up the communication between IO controller PLC and IO devices is established and n...

Страница 16: ...Parameter Range Selection Default Min Max Hz 0 1 200 0 10 0 Analog Output Filter Use The amplifier provides a low pass filter to prevent noise and interferences being overlaid to the analogue output s...

Страница 17: ...Offset The offset value is used to compensate the weight of the measuring roller Unit Parameter Range Selection Default Min Max 8000 8000 0 Gain Use This parameter stores the value determined with th...

Страница 18: ...lue in Newton signed long signed 32 Bit 200 000 000 0 000 N Filtered feedback value re calculated in Newton Feedback Value in Pound signed long signed 32 Bit 200 000 000 0 000 lb Filtered feedback val...

Страница 19: ...0 Request Device Type 0x0 2 R Version unsigned int unsigne d 16 Bit 1 to 10000 0 00 Request Firmware Version 0x1 0 R W Low pass filter feedback active PROFIN ET byte unsigne d 8 Bit 0 bis 1 0 Switch f...

Страница 20: ...ow pass filter feedback PROFINET unsigned int unsigne d 16 Bit 1 to 2000 0 0 Hz Cut off frequency of PROFINET filter 0x2 5 R W Cut off frequency low pass filter Analogue output unsigned int unsigne d...

Страница 21: ...odel of the controller The controller will execute the read write instruction via PROFINET and send back the status or the data to the computer 7 2 Services and Protocols The EMGZ490 is certified in a...

Страница 22: ...ption between EMGZ490 and PLC The interruption usually lasts between 3 5 seconds Afterwards a new connection is automatically established again 7 4 Functional Blocks Example Three functional blocks FB...

Страница 23: ...were defined DB1 FMS_ACYCLIC_DB Name Type Start Value Comment Index INT 0 FMS specific index VALUE_READ DINT L 0 Input value VALUE_WRITE DINT L 0 Output value DB2 FMS_TRIGGER_DB Name Type Start Value...

Страница 24: ...an application in the controller Using a one time write trigger FMS_TRIGGER_DB READ_TRIGGER a read command can be forced The trigger is automatically reset If the read command was successful the data...

Страница 25: ...n t interfere with each other a new module in the PROFINET IO device configuration was introduced The graph Fig 8 shows the actual module configuration Fig 8 Module configuration EMGZ E490014 Each slo...

Страница 26: ...appear see Fig 10 Caution The web interface may only be used in the test mode since data communication can interfere with the PLC Moreover one link can be established via a web browser at the same tim...

Страница 27: ...rrent Reading E490016 The web page Current Reading shows the actual values of the amplifier The first line Tension shows the tension at the input in the configured unit of measure In the second line O...

Страница 28: ...5 28 8 2 Parameter Setting Fig 12 Parameter list E490017 The web page Parameters provides the possibility to configure the amplifier via the web interface It is to notice that in a PROFINET environmen...

Страница 29: ...owser For gain adjustment the page Offset Calibration is available This page offset compensation and calibration These functions are also available via the PLC If the offset and gain values known they...

Страница 30: ...0 15 12 2015 30 8 4 Ethernet Settings This page displays the actual TCP IP configuration It cannot be changed via the web interface Changes to this configuration have to be done over the PLC Fig 14 Et...

Страница 31: ...ration Manual EMGZ 490 15 12 2015 31 8 5 System Settings Via the page System Settings the internal firmware version is displayed Furthermore a new firmware can be up loaded Fig 15 System Settings E490...

Страница 32: ...is available in two different housing options Fig 16 Outline Drawing EMGZ309 R rail mount housing E490002e Ethernet connection RJ 45 connector 108 2 4 26 107 3 4 2 65 2 56 Fig 16 Outline Drawing EMGZ...

Страница 33: ...Operation Manual EMGZ 490 15 12 2015 33 Fig 17 Outline Drawing EMGZ309 W wall mount housing E490003e Ethernet connection M12 4 pole D coded...

Страница 34: ...ng of alarm information LLDP Link Layer Discovery Protocol for proximity detection DCP Discovery and Configuration Protocol request and config of IP Device name etc over network layer SNMP Simple Netw...

Страница 35: ...12 Bit Analogue Output 0 10V min 1 2k Resolution of D A Converter 0 4096 12 Bit Measuring Error 0 05 FS Cycle for Data Exchange 1ms Interface Connectors EMGZ 490 R 2x RJ 45 EMGZ 490 W 2x M12 4 pole D...

Страница 36: ...com FMS Italy Via Baranzate 67 I 20026 Novate Milanese Tel 39 02 39487035 Fax 39 02 39487035 fmsit fms technology com FMS USA Inc 2155 Stonington Ave Suite 119 Hoffman Estates IL 60169 USA Tel 1 847 5...

Отзывы: