Victoreen 190
Operators Manual
3-10
XX is 24, 12, 6, or 3 seconds, corresponding to the newly selected response time.
The
Resp Time
button may be pressed as necessary to scroll through and select the desired response
time. The
Resp Time
button may be disabled via the Model 190-1A Communicator.
The response time displayed is for the lowest
range. Refer to Specifications in Section 1 -
General Information for corresponding response
times on other ranges.
3.13 Resetting the Integrate Value (Resp Time)
If the instrument is in the Integrate Mode and the reset integrator is enabled via the optional Model 190-
1A Communicator, the integrate value may be reset by pressing and holding the
Resp Time
button for
three seconds. The integrate values will reset to zero; the response time will not change. The integrate
reset function may be disabled via the Model 190-1A Communicator.
3.14 Changing the Operation Mode (Mode)
The Model 190 factory default Operational Mode is C/MIN. Using the optional Model 190-1A Infra-red
communicator, default settings can be changed/programmed into the Model 190. Features and push
buttons can also be locked-out to set up the Model 190 in a user-defined mode of operation. The
Operational Mode may be changed by pressing the front panel
Mode
push-button. The next available
mode will then be displayed corresponding to the newly selected Mode. The
Mode
button may be
pressed as necessary to scroll through and select the desired Mode, and includes the following:
1. COUNTS/MIN
MODE
2. COUNTS/SEC
MODE
3. mR/H
MODE
4. Bq/cm
2
MODE
5. μCi/cm
2
MODE
6. DPM
MODE
4. Sv/H
MODE
5. Gy/H
MODE
If the probe has not been initialized and calibrated
(either at the factory or via the operational infrared
communicator), the CPM and CPS modes are the
only modes available. Also, the mCi/sqcm,
Bq/sqcm, and dpm modes require an additional
special calibration performed with the probe. Refer
to section 5 - Calibration for further details.
NOTE
NOTE
Содержание Victoreen 190
Страница 4: ...Blank page...
Страница 12: ...Victoreen 190 Operators Manual Blank page...
Страница 20: ...Victoreen 190 Operators Manual 3 8 Figure 3 5 Probe Adapter Adjustments...
Страница 30: ...Victoreen 190 Operators Manual Blank page...
Страница 32: ...Victoreen 190 Operators Manual Blank page...
Страница 36: ...Victoreen 190 Operators Manual Blank page...