Victoreen 190 

Operators Manual 

 

  3-4 

 
 

 
 

We recommend that the probe be secured to the 
mating probe adapter receptacle using heat shrink 
tubing

.

 

 

4.  Secure the Philips screw located on the top panel of the adapter. 

 

3.6 Power

 

The Model 190 is powered using four 9-volt alkaline batteries. The batteries (four) are housed in a 
separate compartment accessed from the instrument back panel.  All four 9-volt alkaline batteries are 
required to operate the unit for approximately 200 hours.  Visual and audible low battery indications are 
provided whenever battery power is low (four minute intervals for the audible and continuous for the visual 
after a battery test). 

 

3.7 Readout 

Measurement results are indicated on the front panel LCD.  The display includes an analog bargraph, 
scale multiplier, detector range bargraph, and a 16 character alphanumeric display (refer to Figure 3-3).  

Each of the 51 elements in the bargraph represents 2% of the scale span.  The analog bar graph is 
divided into 5 sections with scale notations of 0, 2, 4, 6, 8, and 10.  The scale sections are marked by 
longer bars.  The bars are arranged in an arc with a radius of 3.75 inches (9.5 cm.).  Scale length is 2 
inches (5 cm.). 

The 51 element analog bargraph, located near the top of the display, provides real time response.  It is 
designed to emulate an analog meter movement, making it especially useful in survey applications.  The 
16 character alphanumeric display at the bottom of the LCD displays units, an average digitized value of 
the reading, error messages, integrated value (with elapsed time), operational parameters (e.g., response 
time), low battery indication, probe type and serial number, and data logging information. 

A scale multiplier, located below the bargraph relates the analog bargraph full scale to the units displayed. 

The detector range bargraph consists of five horizontal bars that indicate the decade of the dynamic 
range in cpm of the detector in use.  Most detectors will cover a dynamic counting range of five decades.  
A few detectors (e.g., those with short dead times) can reach the sixth decade of operation.  The sixth 
decade is indicated by all five bars being lit. 

 
 
 

The bargraph may be disabled using the Model 
190-1A Communicator. 

 

In the Alarm condition, the entire display blinks on and off with a period of two seconds

 

NOTE

NOTE

Содержание Victoreen 190

Страница 1: ...190 Survey and Count Rate Meter Operators Manual March 2005 Manual No 190001 Rev 8 2004 2005 Fluke Corporation All rights reserved Printed in U S A All product names are trademarks of their respective...

Страница 2: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 120 Andrews Road Hicksville New York 11801 516 870 0100 www flukebiomedical com rms...

Страница 3: ...Internal Controls 3 4 3 9 High Voltage Adjustment 3 6 3 10 Detector Sensitivity Adjustment 3 8 3 11 Jam Current Adjustment 3 8 3 12 Changing the Response Time RESP TIME 3 8 3 13 Resetting the Integra...

Страница 4: ...Blank page...

Страница 5: ...t indicate the quantity of the dynamic range of the detector in use A 16 character alphanumeric display that 1 provides a digitized average of the bargraph value or integrated reading with elapsed tim...

Страница 6: ...ecifications Radiation Detected Alpha Beta Gamma or X ray depending on detector used Operating Rang Rate Mode CPM 0 100 0 1000 0 10 000 0 100 000 0 1 000 000 and 0 10 000 000 CPS 0 1 0 10 0 100 0 1000...

Страница 7: ...5 6 0 10M 0 7 12 second response time selected Scale Range CPM Response seconds 1 0 100 12 2 0 1000 6 3 0 10K 3 4 0 100K 1 5 5 0 1M 1 5 6 0 10M 0 7 6 second response time selected Scale Range CPM Resp...

Страница 8: ...dio on off and acknowledges alarm conditions ON OFF Turns instrument power on off when permitted by the microprocessor i e when data logging is inactive or all logged data has been read by a PC Intern...

Страница 9: ...instrument The pulse shaping circuit in the module accepts a wide range of detector input amplitudes and pulse decay times Communication with the Model 190 is accomplished through an infrared two way...

Страница 10: ...suitable calibration constant is used Above 384 000 CPM the counts in a time slot are counted and the value multiplied by the appropriate factor to determine the actual count rate which may cause a re...

Страница 11: ...inimum of one second The amount of time remaining before the unit stops operating depends on previous use of the instrument If the instrument is used for longer continuous periods of time there will b...

Страница 12: ...Victoreen 190 Operators Manual Blank page...

Страница 13: ...osive materials fluctuations in temperature and humidity and vibration and shock 3 3 Battery Installation Replacement Four 9 volt alkaline batteries are supplied with the Model 190 The battery compart...

Страница 14: ...batteries are installed 3 4 Probe Holder Installation Extreme caution should be used when connecting the probe to the meter Improper connection may result in injury damage to the instrument or damage...

Страница 15: ...atibility was requested on the purchase order If a probe was ordered with the instrument it is connected to the probe adapter Use the following procedure to replace a probe and adapter 1 Loosen the ca...

Страница 16: ...are marked by longer bars The bars are arranged in an arc with a radius of 3 75 inches 9 5 cm Scale length is 2 inches 5 cm The 51 element analog bargraph located near the top of the display provides...

Страница 17: ...ewing and Fluke Biomedical does not recommend that it be further adjusted However if the display contrast requires adjustment turn RP1 to vary the angle of maximum contrast of the 16 character display...

Страница 18: ...adapter 2 Unplug the adapter from the unit 3 Remove the flat head screw from the bottom panel of the adapter 4 Remove the top panel of the adapter to gain access to the inner components There is a pot...

Страница 19: ...screws located on the top panel of the adapter 3 9 HIGH VOLTAGE ADJUSTMENT Determine which module you have surface mount or non surface mount before continuing and ensure that the correct figure is r...

Страница 20: ...Victoreen 190 Operators Manual 3 8 Figure 3 5 Probe Adapter Adjustments...

Страница 21: ...current 1 Adjust RP1 fully counter clockwise 2 Bring the detector to a high count rate until the reading begins to drop with increasing intensity 3 Adjust RP1 clockwise until the jam message appears o...

Страница 22: ...or 3 14 Changing the Operation Mode Mode The Model 190 factory default Operational Mode is C MIN Using the optional Model 190 1A Infra red communicator default settings can be changed programmed into...

Страница 23: ...be activated through the optional Communicator If the Log Mode has not been activated a NOT IN LOG MODE message will be displayed when the Log button is pressed When in the data logging mode the instr...

Страница 24: ...ode is active Log changes the label Rate Integ toggles the alphanumeric display between the digitized rate value and the integrate value Resp Time changes the response time on the first range or reset...

Страница 25: ...s not been initialized and calibrated either at the factory or via the optional infrared communicator the CPM and CPS modes are the only modes available In addition the mCi sqcm Bq sqcm and dpm modes...

Страница 26: ...he detector The clicks from the speaker have a one to one correspondence with counts from the detector up to 3200 cpm For counts above 3200 cpm there is a transition from clicks to a steady tone g Pre...

Страница 27: ...edical at 440 248 9300 for further instructions 2 If a problem occurs with the printed circuit board case or audible alarm transducer the suspect part is easily replaceable Refer to Table 4 1 for a li...

Страница 28: ...stener 2 190031 Case Bottom 190023 Battery Cover 190035 Battery Cover Gasket 149036 Transducer Screen 67 115 Connector 190015 Keypad Assembly 5 1154 Phillips Screw 5 6 19 X 1 25 PH 5 1153 Phillips Scr...

Страница 29: ...ration data stored in the probe adapter may be changed using the optional Communicator P N 190 1A Refer to the instruction manual supplied with the Communicator for further instructions and specific p...

Страница 30: ...Victoreen 190 Operators Manual Blank page...

Страница 31: ...icable to the Model 190 and contain proprietary information If you require these drawings please contact Fluke Biomedical at 440 248 9300 Drawings Drawing Number Description 190013SM Model 190 Schemat...

Страница 32: ...Victoreen 190 Operators Manual Blank page...

Страница 33: ...e of viewing The metal stand is rugged being made of formed and welded steel The Model 190AC can be easily inserted and removed from this stand Dimensions The dimensions of the Model 190AC are the sam...

Страница 34: ...his source will not damage the 190AC or the adapter module The metal stand has holes in the base and back to permit permanent installation if desired Operational Procedure The Model 190AC may be inser...

Страница 35: ...djustments For calibration and adjustments to the Model 190AC refer to the Model 190 and 190AC sections of this manual Replacement Parts Item Part Number Model 190AC 190AC Metal Stand 190232 Ac to Dc...

Страница 36: ...Victoreen 190 Operators Manual Blank page...

Страница 37: ...uded Item Qty Adapter bracket assembly 190235 1 6 32 Machine Screws 3 6 Split Washers 3 6 32 Hex Nuts 3 3 Align the three holes on the side of the adapter bracket assembly 190235 with the three holes...

Страница 38: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 120 Andrews Road Hicksville New York 11801 516 870 0100 www flukebiomedical com rms...

Отзывы: