2-26
VIBXPER
T II Balancing 12.2019
Setup parameters
To execute the balancing procedure VIBXPERT requires different set-
tings, which are collected in ‘setup menus’:
- Machine setup (see below)
- Measurement setup (see p. 2-28)
- Transducer setup (see VIBXPERT II operating manual, page 2-20)
These setups are configured prior to a balancing procedure. Dur-
ing balancing, only certain parameters from the machine setup can
be changed. These parameters can be found in the Tools menu (p.
2-19f.).
Machine setup
A 'machine setup' contains the machine data required for balancing.
VIBXPERT uses these data to calculate a suitable trial weight and the
residual imbalance, the centrifugal force and the attained balancing
quality. In order to edit or create a machine setup, proceed as follows:
• Mark a rotor type in the selection screen (see p. 2-11).
• Press the MENU button and click on "Task manager".
• Select a user-defined task.
• Mark the "Machine setup" field.
• Press the MENU button and click on "New" or on "Edit":
The following parameters can be set:
p
lanes
:
One / Two
. Number of balancing planes; cannot be changed
because the rotor type is selected in the selection screen.
c
orrection
MoDe
*
:
Free / fixed location / fixed weight / meas. tape
;
takes into account restrictions on the machine (fixed locations, e.g.
for ventilator blades) or local circumstances (tape measure).
Machine setup
*This parameter can be changed in each
balancing run (MENU - Tools)
Parameters in the machine setup